Chapter 3 Cleaning, Disinfection and Sterilization Procedures
57
EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL
Figure 3.32
4.
Seal the instrument in a package appropriate for ETO gas sterilization
according to your hospital’s protocol.
5.
Sterilize the package according to the recommended ETO gas exposure
parameters described in Section 2.5, “ETO gas sterilization” and the
sterilizer manufacturer’s instructions.
6.
Aerate the components following the minimum aeration parameters
specified in Section 2.5, “ETO gas sterilization”.
7.
Store the components following the instructions given in Chapter 5, “Storage
and Disposal”.
CF-Q160ZL/I, CF-Q160DL/I