Chapter 3 Cleaning, Disinfection and Sterilization Procedures
59
EVIS EXERA GIF/CF/PCF TYPE 160 Series REPROCESSING MANUAL
Reusable parts which are normally reprocessed separately
from the endoscope
Manual cleaning
• Make sure that the items immersed in detergent solution do
not contact one another.
• Make sure not to scratch the seals on the air/water valve and
AW channel cleaning adapter with brushes, etc.
1.
Fill a basin with clean water and low-foaming detergent solution at the
temperature and concentration recommended by the detergent
manufacturer. Use a basin which is deep enough to allow all equipment to
be completely immersed.
2.
Immerse all equipment in the detergent solution. For the biopsy valve,
remove the cap from the body. Using a clean, soft brush or lint-free cloth,
meticulously clean all external surfaces in detergent solution.
3.
Clean the bristles of the channel cleaning brushes thoroughly while the
brushes are immersed.
4.
While immersed, depress and release the pistons of the air/water valve,
suction valve and AW channel cleaning adapter.
5.
Using the channel cleaning brush, thoroughly brush the openings of the
suction valve, air/water valve and biopsy valve until no debris can be seen.
Clean the bristles of the brush while it is immersed in the detergent solution.
6.
To clean the suction cleaning adapter, attach the 30 cm
3
(30 ml) syringe
and flush the tube thoroughly with detergent solution.
• Air/water valve
(MH-438)
• Suction valve
(MH-443)
• Biopsy valve
(MB-358)
• Mouthpiece
(MA-474, MB-142)
• AW channel cleaning adapter
(MH-948)
• Suction cleaning adapter
(MH-856)
• Channel cleaning brush
(BW-20T)
• Channel cleaning brush
(BW-17K for GIF-XTQ160 only)
• Channel-opening cleaning brush
(MH-507)
• Auxiliary water tube
(MAJ-855 for endoscopes with
auxiliary water feeding only)