5 The recommended GATE position
IMPULSE display
IMPULSE
mode applies an envelope filter over the RF signal. (The name “
IMPULSE
”
originates from earlier BondMaster products.)
6.
If preferred, set the
RF DISPLAY
(E key) to
IMPULSE
page 127). However, it is recommended to use the
RF DISPLAY
mode because it
allows easier interpretation of each signal oscillation.
NOTE