DMTA-10045-01EN, Rev. E, August 2016
Applications
169
Figure 6
‑
63 The signal for the smaller defect
5.
Hold the probe over a defect-free section, and then press
GOOD PART
(E key)
[see Figure 6-64 on page 169].
Figure 6
‑
64 The signal for the defect
‑
free zone
6.
If necessary, select the best operating frequency for your needs by adjusting the
FREQ
(D key) using the knob (see Figure 6-65 on page 170).
In many situations, the BondMaster 600 automatically picks the best operating
frequency. However, for some more complex or noisier applications, it is
recommended to manually select the frequency.
Содержание BondMaster 600
Страница 8: ...DMTA 10045 01EN Rev E August 2016 Table of Contents viii ...
Страница 10: ...DMTA 10045 01EN Rev E August 2016 List of Abbreviations x ...
Страница 26: ...DMTA 10045 01EN Rev E August 2016 Important Information Please Read Before Use 16 ...
Страница 28: ...DMTA 10045 01EN Rev E August 2016 Introduction 18 ...
Страница 32: ...DMTA 10045 01EN Rev E August 2016 Chapter 1 22 ...
Страница 72: ...DMTA 10045 01EN Rev E August 2016 Chapter 3 62 ...
Страница 78: ...DMTA 10045 01EN Rev E August 2016 Chapter 4 68 Figure 4 2 Crosshairs and the NULL point ...
Страница 130: ...DMTA 10045 01EN Rev E August 2016 Chapter 5 120 ...
Страница 141: ...DMTA 10045 01EN Rev E August 2016 Applications 131 Figure 6 13 RUN 3 XY FLY DOT Figure 6 14 RUN 4 XY SCAN ...
Страница 216: ...DMTA 10045 01EN Rev E August 2016 Chapter 7 206 ...
Страница 230: ...DMTA 10045 01EN Rev E August 2016 Appendix A 220 ...
Страница 234: ...DMTA 10045 01EN Rev E August 2016 Appendix B 224 ...
Страница 240: ...DMTA 10045 01EN Rev E August 2016 List of Figures 230 ...
Страница 242: ...DMTA 10045 01EN Rev E August 2016 List of Tables 232 ...