DMTA-10045-01EN, Rev. E, August 2016
Chapter 3
58
If the BondMaster 600 is being used in noisy conditions, an external horn can be used.
The horn connects to the I/O connector on the back of the BondMaster 600 and
increases the audible alarm’s sound output to 70 dB (see Table 8 on page 209 for the
part specification).
3.4
Displaying Real-Time Readings
The real-time readings bar displays user-configurable readings (measurements) [see
Figure 3-5 on page 55]. It can display a maximum of two real-time readings from
among the available choices—the number of available choices depends on the
selected operating mode. The real-time readings bar can be set to display one or two
readings, or it can be disabled.
The real-time readings bar can display the following readings (see Figure 3-7 on
page 59 and Figure 3-8 on page 59):
•
LIVE AMPL
— Maximum distance between current XY (horizontal, vertical) dot
position (swept mode excluded) and null position.
•
LIVE VERT
— Maximum distance between current vertical (Y) dot position
(swept mode excluded) and null position.
•
LIVE HORZ
— Maximum distance between current horizontal (X) dot position
(swept mode excluded) and null position.
•
LIVE ANGL
— Angle of current XY position with respect to the null point (swept
mode excluded).
•
AMPLITUDE P
‑
P
— Volts peak-to-peak or largest signal (swept mode only).
NOTE
Содержание BondMaster 600
Страница 8: ...DMTA 10045 01EN Rev E August 2016 Table of Contents viii ...
Страница 10: ...DMTA 10045 01EN Rev E August 2016 List of Abbreviations x ...
Страница 26: ...DMTA 10045 01EN Rev E August 2016 Important Information Please Read Before Use 16 ...
Страница 28: ...DMTA 10045 01EN Rev E August 2016 Introduction 18 ...
Страница 32: ...DMTA 10045 01EN Rev E August 2016 Chapter 1 22 ...
Страница 72: ...DMTA 10045 01EN Rev E August 2016 Chapter 3 62 ...
Страница 78: ...DMTA 10045 01EN Rev E August 2016 Chapter 4 68 Figure 4 2 Crosshairs and the NULL point ...
Страница 130: ...DMTA 10045 01EN Rev E August 2016 Chapter 5 120 ...
Страница 141: ...DMTA 10045 01EN Rev E August 2016 Applications 131 Figure 6 13 RUN 3 XY FLY DOT Figure 6 14 RUN 4 XY SCAN ...
Страница 216: ...DMTA 10045 01EN Rev E August 2016 Chapter 7 206 ...
Страница 230: ...DMTA 10045 01EN Rev E August 2016 Appendix A 220 ...
Страница 234: ...DMTA 10045 01EN Rev E August 2016 Appendix B 224 ...
Страница 240: ...DMTA 10045 01EN Rev E August 2016 List of Figures 230 ...
Страница 242: ...DMTA 10045 01EN Rev E August 2016 List of Tables 232 ...