DMTA-10045-01EN, Rev. E, August 2016
Applications
149
3.
Select the
Metal To Metal Disbonds
application, and then press
to accept
(see Figure 6-38 on page 149).
Figure 6
‑
38 The Metal To Metal Disbonds application
4.
If the calibration menu does not appear automatically, press and hold the CAL
NULL key (
).
5.
Hold the probe in the air. The BondMaster 600 should automatically pick the best
operating frequency for the probe. When in doubt, press
CAL
(C key) or change
the
FREQ
(D key) using the knob.
6.
Press
DONE
(E key) [see Figure 6-39 on page 149].
Figure 6
‑
39 The CAL screen
Содержание BondMaster 600
Страница 8: ...DMTA 10045 01EN Rev E August 2016 Table of Contents viii ...
Страница 10: ...DMTA 10045 01EN Rev E August 2016 List of Abbreviations x ...
Страница 26: ...DMTA 10045 01EN Rev E August 2016 Important Information Please Read Before Use 16 ...
Страница 28: ...DMTA 10045 01EN Rev E August 2016 Introduction 18 ...
Страница 32: ...DMTA 10045 01EN Rev E August 2016 Chapter 1 22 ...
Страница 72: ...DMTA 10045 01EN Rev E August 2016 Chapter 3 62 ...
Страница 78: ...DMTA 10045 01EN Rev E August 2016 Chapter 4 68 Figure 4 2 Crosshairs and the NULL point ...
Страница 130: ...DMTA 10045 01EN Rev E August 2016 Chapter 5 120 ...
Страница 141: ...DMTA 10045 01EN Rev E August 2016 Applications 131 Figure 6 13 RUN 3 XY FLY DOT Figure 6 14 RUN 4 XY SCAN ...
Страница 216: ...DMTA 10045 01EN Rev E August 2016 Chapter 7 206 ...
Страница 230: ...DMTA 10045 01EN Rev E August 2016 Appendix A 220 ...
Страница 234: ...DMTA 10045 01EN Rev E August 2016 Appendix B 224 ...
Страница 240: ...DMTA 10045 01EN Rev E August 2016 List of Figures 230 ...
Страница 242: ...DMTA 10045 01EN Rev E August 2016 List of Tables 232 ...