10-015490-01EN, Rev. 2, June 2020
Appendix C
84
Figure C-5 Background subtraction menu
2.
Without changing any of the parameters, click
Do it!
3.
Click
Yes
in the confirmation dialog box.
4.
On the XPowder tool bar, click
Advanced searching
(
).
The advanced searching window containing a diffractogram appears.
5.
In the diffractogram, click within the (blue) background subtraction area to
decrease the amount of background subtraction (see Figure C-6 on page 84).
The purpose of decreasing or increasing the amount of background subtraction is
to eliminate as much noise as possible while preserving as many peaks as
possible.
Figure C-6 Advanced searching windows
Background subtraction set too high — Some peaks
are eliminated.
Background subtraction decreased — Most peaks
are preserved.