
10-015490-01EN, Rev. 2, June 2020
Set Up and Operation
47
4.
Press the Check button (
) to start the test.
When the test begins, the BTX III analyzer gives the data set a name based on a
sequential number. The red LEDs glow, indicating the application of high voltage
to the X-ray tube. The display shows the voltage and intensity readings, and the
sample carrier emits a high-pitched vibration noise.
To stop testing before all the exposures are taken
Press the Stop button (
) once.
Shaking
X to quit
Initiates vibration of the
external shaker assembly,
assuming it is connected.
Press the Stop button on
the keypad to terminate
external shaking. Adjust
the shaking amplitude
with the left and right
arrows. Save this value
with the Check key.
Do USB
Transfer
Choose
Dataset
(Data set)
Selects the data set that is
transferred to the USB flash
drive.
Shut Down
Turns off the BTX III
analyzer. The system takes
a few minutes to slowly
warm up the detector,
preventing condensation
from forming and avoiding
damage.
Table 8 Running modes
(continued)
Menu level
one
Menu level
two
Level two choice
Comments
Содержание BIX III
Страница 6: ...Table of Contents vi 10 015490 01EN Rev 2 June 2020...
Страница 8: ...10 015490 01EN Rev 2 June 2020 List of Abbreviations viii...
Страница 12: ...10 015490 01EN Rev 2 June 2020 Labels and Symbols 4...
Страница 28: ...10 015490 01EN Rev 2 June 2020 Introduction 20...
Страница 36: ...10 015490 01EN Rev 2 June 2020 Chapter 1 28...
Страница 64: ...10 015490 01EN Rev 2 June 2020 Chapter 4 56...
Страница 82: ...10 015490 01EN Rev 2 June 2020 Chapter 5 74...
Страница 84: ...10 015490 01EN Rev 2 June 2020 Appendix A 76...
Страница 100: ...10 015490 01EN Rev 2 June 2020 List of Tables 92...