43
3.
Press
Factory
button (see pointer 3 in Figure 58 b) to view the factory calibration parameters. Press
Use
factory
to recover and use the factory calibration parameters (see Figure 60 b).
Note
►
T Sensor is factory calibrated at three different temperatures: 35.0°C, 37.0 °C and 40.0°C.
a
b
Figure 60. Settings
–
Calibrations
–
T Sensor
–
Factory (a
–
b).
4.
By pressing
Set
(see pointer 4 in Figure 58 b) UNO-CONTROLLER adjusts the reading of the T Sensor
against the reading of the certified thermometer.
To start the T Sensor calibration, follow the instructions below:
•
Make sure that T Sensor is connected to UNO-CONTROLLER (see paragraph 7.4).
•
Set a water bath to the temperature you use as set point for the heating devices during standard
operation (most of the times it is 37°C);
•
Measure water temperature by immersing the T Sensor and a certified thermometer;
•
Press on the pencil icon
(see Figure 61 a) and insert the reading of the certified thermometer by
pr / - (see pointer 1 in Figure 61 b). Press
OK
to confirm (see pointer 2 in Figure 61 b).
a
b
2
1
Figure 61. Settings
–
Calibrations
–
T Sensor
–
Set (a
–
b).
•
Press on
Set
(see pointer 4 in Figure 58 b) to adjust to reading of the T Sensor against the reading of
the certified thermometer.
Tip
►
Press on icon
to receive instructions on how to calibrate the T Sensor.
Note
►
Okolab recommends using a certified thermometer as external reference.