10.1
CO2-O2 Measurement process: Preliminary Operations
Before starting the CO2-O2 Measurement Process, you have to follow some preliminary operations. See
paragraph 10.1.1 for the CO2-O2 measurement in Diffusion Mode or paragraph 10.1.2 for the CO2-O2
measurement in Aspiration Mode.
10.1.1
Diffusion Mode
–
Preliminary Operations
1.
Connect the
clear connector
of the Tube-A to LEO gas input port labeled
(1 in Figure 46,
Configuration a and in Figure 46, configuration b).
2.
Put the cap on LEO gas input port labeled with icon
(2 in Figure 46, Configuration a and in Figure
46, configuration b)
3.
Connect the
red connector
of the Tube-A to the
green connector
of Tube-B (3 in Figure 46,
Configuration a).
If the sampling gas is humid, connect the
red connector
of Tube-A (3 in Figure 46, Configuration b) and
the
green connector
of Tube-B (4 in Figure 46, Configuration b) to the Moisture Trap. In this case,
operate in
Wet Mode
(see paragraph 10.1.5)
4.
Connect the end of Tube-B to the incubator or to the point where measurement is required (4 in Figure
46, Configuration a and 5 in Figure 46, Configuration b)
5.
Okolab recommends the use of a purge tube attached to LEO gas outlet port (5 in Figure 46,
Configuration a and 6 in Figure 46, Configuration b ). You can use the supplied additional tube as tube
for purging. Use the supplied female luer connector to connect the additional tube to LEO gas outlet
port. The purge tube has to be placed in a well ventilated environment.
Tip
►
Use the fitting tubes and additional tubes in order to allow the compatibility of the supplied tubing
with the sampling gas port of your incubators.
Содержание LEO
Страница 1: ...LEO Handheld Analyzer for IVF Applications Manual IST1446_REV05 SV 3 1 64 0 ...
Страница 2: ...This page was left blank ...
Страница 16: ...12 Figure 13 Pump ...
Страница 84: ...80 Figure 110 T1 Calibration Factory Reset a b b a ...
Страница 88: ...84 f e c a b d 2 1 1 Figure 114 CO2 O2 MODULE insertion ...
Страница 92: ...88 f e c a b d 2 1 1 Figure 118 CO2 O2 MODULE insertion after pump replacement ...