22
9.3.3
Pump
Figure 24 shows the internal pump subpages. By pressing
Pump
icon
(see Figure 24 a), you can read
the remaining lifetime of the pump and its current flow rate.
Press
Advanced
(see Figure 24 b) to set a different flow rate (from 60 to 260 ml/min) (see Figure 24 c).
Tip
►
The default flow rate value is 150 ml/min.
You can also reset the remaining lifetime if a new pump i
s installed by pressing on “
Reset
c). See the paragraph 18 for the pump replacement.
Note
►
The pump lifetime must be reset only if you replace the pump with a new one.
a
b
c
Figure 24. Pump Page (a
–
b - c).
9.3.3.1
Pump Leak Test
You can perform the Leak test to verify the proper functioning of the pump.
a
b
c
Figure 25. Leak test selection (a
–
b
–
c).