92
21.2
CO2-O2-MODULE
CO2 Sensor
–
Technical Data
Sensor type
Non Dispersive InfraRed detector (NDIR)
Measurement Range
0-20%
Accuracy
± (1% of Full Scale +2% Reading) in the range 0-10%
1
± (2% of Full Scale +3% Reading) in the range 11-20%
1
Response Time, T
90
< 35 s
O2 Sensor
–
Technical Data
Sensor type
Fluorescence-based optical sensor
Measurement Range
0-25%
Accuracy
± (1% of Full Scale) in the range 0-10%
1
± (2% of Full Scale +3% Reading) in the range 11-25%
1
Response Time, T
90
< 35 s