When CO2-O2 Logging is completed or if you stop it by clicking on the
icon (see Figure 52 c), the last values
remain on the Homepage for 10 seconds, or until a new measurement occurs.
5.
Press the
View
icon
(see Figure 51 a) to view data stored. More details are reported in chapter
10.1.5
Wet Mode
LEO measures the sampled gas composition at the environmental pressure, temperature and humidity
conditions. If the gas sample is very humid, you must use the moisture trap between the incubator and LEO (see
paragraph 5.2 and paragraph 10.1), the loss of water molecules changes the composition of the gas and has an
effect on the CO2
–
O2 measurement. To address this issue and correct the CO2 and O2 measurement to account
for the incubator real condition, you can use the
Wet Mode
.
The
Wet Mode
is a LEO feature allowing you to set the incubator temperature and relative humidity values.
LEO will correct the CO2 and O2 values by accounting for the real water content of the gas drawn from the
incubator.
In order to set
Wet Mode
, press the
Wet/Dry Mode
button
(see Figure 53 a). A pop-up dialog box
appears. Insert the temperature and relative humidity values of the Wet Incubator and then press
Save
, as shown
in Figure 53 b.
Figure 53. How to select Wet Mode (a
–
b).
a
b
Содержание LEO
Страница 1: ...LEO Handheld Analyzer for IVF Applications Manual IST1446_REV05 SV 3 1 64 0 ...
Страница 2: ...This page was left blank ...
Страница 16: ...12 Figure 13 Pump ...
Страница 84: ...80 Figure 110 T1 Calibration Factory Reset a b b a ...
Страница 88: ...84 f e c a b d 2 1 1 Figure 114 CO2 O2 MODULE insertion ...
Страница 92: ...88 f e c a b d 2 1 1 Figure 118 CO2 O2 MODULE insertion after pump replacement ...