Do not use a different power supply to recharge LEO. Do not attach a different USB cable to the
power adapter.
5.2
Soda Lime Kit
The Soda Lime Kit (D in Figure 3) is used for the
CO2 Zero Reset
procedure with ambient air (see paragraph
13.4.1). It consists of a Soda Lime filter (1 in Figure 6) generally used to remove CO2 from the ambient air.
The Soda Lime Kit must be connected to Tube-A to start the
CO2 Zero Reset
procedure using ambient air.
The
clear connector
(2 in
Figure 6) of Soda Lime Kit has to be connected to the
red connector
of Tube-A (1 in
Note
►
The Soda Lime kit has to be replaced when the Soda Lime filter (1 in Figure 6) changes color turning
purple. Contact Okolab to purchase a new Soda Lime Kit.
Note
►
Remove the protection cap from the Soda Lime filter before using the Soda Lime Kit.
Figure 6. Soda Lime Kit.
5.3
Moisture Trap
The Moisture Trap (E in Figure 3) is used for humid gas measurements.
Figure 7 shows the Moisture Trap components:
1.
A condensation trap allowing for the reduction of sampled gas temperature and the collection of
condensed water from the sampled gas (4 in Figure 7). The condensation trap top (3 in Figure 7)
allowing reducing the sampling gas time during the Single Point measurement must be removed when
Gas Logging occurs.
2.
A Selemion Tube that reduces the relative humidity of sampled gases to a value lower than 40% (5 in
The
connector marked in green
is the Gas Inlet port of the Moisture Trap (1 in Figure 7). Please use
Tube-B to connect the sampling Gas Port of the incubator to Gas Inlet port of the Moisture Trap.
The
connector marked in red
is the Gas Outlet port for the Moisture Trap (2 in Figure 7). Please use
the Tube-A to connect the Gas Outlet port of the Moisture Trap to the Gas Inlet port of LEO (see Configuration
b in Figure 46 for the measurements in Diffusion Mode or Configuration b in Figure 48 for the measurements
in Aspiration Mode).
1
2
Содержание LEO
Страница 1: ...LEO Handheld Analyzer for IVF Applications Manual IST1446_REV05 SV 3 1 64 0 ...
Страница 2: ...This page was left blank ...
Страница 16: ...12 Figure 13 Pump ...
Страница 84: ...80 Figure 110 T1 Calibration Factory Reset a b b a ...
Страница 88: ...84 f e c a b d 2 1 1 Figure 114 CO2 O2 MODULE insertion ...
Страница 92: ...88 f e c a b d 2 1 1 Figure 118 CO2 O2 MODULE insertion after pump replacement ...