Press on
CO2 Zero Reset
icon
(see Figure 91) to correct the zero of the CO2 sensor, the pop-up in
Figure 91 b appears.
You can decide to use
Air
or
N2
for the
CO2 Zero Reset
procedure.
Tip
►
Okolab suggests to start the CO2 Zero Reset using N2, use Air only if N2 is not available.
a
b
Figure 91. CO2 Zero Reset (a
–
b).
13.4.1
CO2 Zero Reset with Air
Press on
Air
to start the
CO2 Zero Reset if you are
using air. See Figure 92 a.
a
b
c
Figure 92. CO2 Zero Reset with Air (a
–
b - c).
Pressing on
Air,
LEO reminds you to connect the Soda Lime kit, Figure 92 b.
To connect the Soda Lime Kit to LEO follow the instructions below:
1.
Connect the
clear connector
of Tube-A to LEO gas inlet port labeled with icon
(1 in Figure 94).
2.
Put the cap on LEO gas inlet port labeled (2 in Figure 94).