37
6.
Press the
Pump/Diffusion Mode
button
(see Figure 47 a) to select the Diffusion Mode (see Figure 47 b).
Once you have followed the step mentioned above, LEO is ready to be used in Diffusion Mode. See
paragraph 10.1.3 to start a Single Point Measurement or paragraph 10.1.4 to start a Gas Logging.
Figure 47. How to select the Diffusion Mode (a
–
b).
10.1.2
Aspiration Mode
–
Preliminary Operations
1.
Connect the
clear connector
of the Tube-A to LEO gas input port labeled
(1 in Figure 48,
Configuration a and in Figure 48, Configuration b).
2.
Put the cap on LEO gas input port labeled with icon (2 in Figure 48, Configuration a in Figure 48,
Configuration b).
3.
Connect the
red connector
of the Tube-A to the
green connector
of Tube-B (3 in Figure 48,
Configuration a).
If the sampling gas is humid, connect the
red connector
of Tube-A (3 in Figure 48, Configuration b) and
the
green connector
of Tube-B (4 in Figure 48, Configuration b) to the Moisture Trap. In this case,
operate in
Wet Mode
(see paragraph 10.1.5)
4.
Connect the end of Tube-B to the incubator or to the point where measurement is required (4 in Figure
48, Configuration a and 5 in Figure 48, Configuration b)
5.
Okolab recommends the use of a purge tube attached to LEO gas outlet port (5 in Figure 48,
Configuration a and 6 in Figure 48, Configuration b ). You can use the supplied additional tube as tube
for purging. Use the supplied female luer connector to connect the additional tube to LEO gas outlet
port. The purge tube has to be placed in a well ventilated environment.
Tip
►
Use the fitting tubes and additional tubes in order to allow the compatibility of the supplied tubing
with the sampling gas port of your incubators.
a
b
Содержание LEO
Страница 1: ...LEO Handheld Analyzer for IVF Applications Manual IST1446_REV05 SV 3 1 64 0 ...
Страница 2: ...This page was left blank ...
Страница 16: ...12 Figure 13 Pump ...
Страница 84: ...80 Figure 110 T1 Calibration Factory Reset a b b a ...
Страница 88: ...84 f e c a b d 2 1 1 Figure 114 CO2 O2 MODULE insertion ...
Страница 92: ...88 f e c a b d 2 1 1 Figure 118 CO2 O2 MODULE insertion after pump replacement ...