27
Calibration offset
Calibration ID
Calibration Date and
Time
Username
Total Calibration events
Start time of the logging
Figure 34. Logging of calibration data
You can recall any calibration by pressing on the corresponding row, as shown in Figure 35(a) and you
will be asked if you want to apply the stored offset on the selected channel, as shown in Figure 35(b).
a
b
Figure 35. How to apply a previous calibration
Press on
To USB
if you want to download the calibration data on the USB drive or on
Erase
if you want to
delete them.
Tip 13
►
We recommend protecting this feature with a Password (see 8.2.6).
8.2.2.2
T Sensor Calibration
The T Sensor of H401-T-CONTROLLER is calibrated in Okolab at three different temperatures, i.e. 35.0,
37.0 and 40.0°C. To view the factory calibration data, press on the T sensor calibration icon
as shown in
Figure 36(a) and then on the tab labeled
Factory
as shown in
Figure 36(b); the page with the calibration data will
open, as shown in Figure 37.