Figure 29. Heating device manual calibration. How to start manual calibration
Figure 30. Manual Calibration. How to introduce the offset value
a
b
Figure 31. Username input. (a) Selection of the keyboard; (b) Operator name Input
Tip 12
►
Okolab recommends starting with an offset value corresponding to the difference between the set
point temperature and the actual temperature measured by the T Sensor. In Figure 30, the desired
temperature is 37.0 °C while the sensor reading states 36.8 °C, hence, in this case the offset to be applied is
0.2 °C.
5
Wait until the temperature measured by the T Sensor is stable.
6
If the temperature measured by the T Sensor is close to the set point, you can consider the Manual
Calibration as completed. Otherwise, repeat the steps above until the temperature measured by the T
Sensor reaches the set point temperature.