33
Note 20
► The
transient time range is 30-240 minutes.
Note 21
►
The transient time should be twice time the typical time to get the thermal plate in temperature
at the set point. So if the thermal plate reaches the set point temperature stably after 10 minutes, let me
suggest to set the transient time to 20 minutes.
Tip 19
►
After any change in the set point value the device enters into a transient regime.
4
Flag the tab labeled
Buzzer
if you want the alarm to be acoustical, additionally to being visual, see Figure
48(b)
5
Flag the tab labeled
Enabled
to enable the Alarms logic, see Figure 48(b)
6
To disable the visual and acoustical alarms, remove the flag from the tab labeled
Enabled
. In this case the
Alarms page becomes as shown in Figure 49
Figure 49. Alarms page when the alarms are disabled
Tip 20
►
When disabling the alarms, H401-T-CONTROLLER goes into alarm condition only when the
status LED becomes red (for example is a device is accidentally disconnected)
Note 22
►
By Pressing on Log (see Figure 50), you can view all the times that the status LED of the two
devices have changed color (see Figure 51).
Figure 50. How to enter in alarms logging page