28
a
b
Figure 36. Calibration data. (a) T sensor calibration page selection; (b) T sensor calibration page
Figure 37. Free sensor factory offset
To re-calibrate the T sensor, follow the instructions below.
1
Set a water bath to the temperature you use as set point for the heating devices during standard operation
(most of the times it is 37°C).
2
Measure water temperature by immersing the T-Sensor and a Certified Thermometer.
3
Type in the reading of the certified thermometer by pressing on the pencil
, as shown in Figure 38(a)
4
Press on
Set
to adjust to reading of the T-Sensor against the reading of the certified thermometer, as
shown in Figure 38(b).
a
b
Figure 38. How to re-calibrate the T sensor
Tip 14
►
Press on icon
to receive instructions on how to calibrate the T-Sensor.