CHAPTER 3 MAINTENANCE PROCEDURES
Ohaus Corporation www.ohaus.com
3-36
Pioneer
™ Balances Service Manual
3.3.22 Internal Calibration Testing
This section describes how to test the internal calibration in a Block Style Load Cell. There can
be several reasons that internal calibration does not function, including a broken or jammed
InCal Mechanism, a defective InCal Motor, or a defective Main PC Board. To test the internal
calibration, proceed as follows:
1. Connect an RS232 cable and a computer to the balance.
2. Follow instructions in Appendix C, Section C.5.1, Incal Weight Mechanism Testing.
3. Observe the motion of the InCal Weight when you press
Raise
or
Lower
.
If the weight does not move, replace the InCal Motor. (See Section 3.3.21.)
4. If the operation is not smooth, replace the entire InCal Mechanism. (See Section
3.3.22.)
3.3.23 InCal Motor Removal & Replacement
1.
Remove the balance’s Housing and Load Cell, as described in earlier sections.
2.
Remove the two screws holding the InCal Motor and InCal Motor Shield in place
from the InCal Mechanism. (See Figures 3-99 and 3-100.)
Figure 3-99. InCal Motor.
Figure 3-100. InCal Motor Removal.
3.
With the Incal Weight in upper-most
position, the Motor Coupler should
be in the position shown in
Figure 3-101.
Figure 3-101. Motor Coupler.
Screws holding
InCal Motor
InCal Motor
Art
InCal
Mechanism
Art
InCal Motor
Shield
Art
Mounting
Screws
Art
InCal Weight
Motor
Coupler
Содержание Pioneer PA64
Страница 1: ...SERVICE MANUAL Ohaus Corporation 19A Chapin Road P O Box 2033 Pine Brook NJ 07058 2033 973 377 9000 ...
Страница 2: ......
Страница 10: ...TABLE OF CONTENTS Pioneer Balances Service Manual viii Ohaus Corporation www ohaus com ...
Страница 106: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 18 Pioneer Balances Service Manual ...
Страница 115: ......