APPENDIX A STANDARD CALIBRATION
Ohaus Corporation www.ohaus.com
A-2
Pioneer
™ Balance Service Manual
A.3 Linearity Calibration
With the balance on, press and release
Unit/Menu.
until
CAL
appears. Press
Yes
(the
On/Off
button).
.CAL.
SPAN
appears. Press No.
Linear
appears. Press
Yes
.
(If
PUSH
flashes, place maximum weight on the Pan.)
Linear
After a few seconds
Clear Pan
flashes.
Remove the mass from the Pan.
Clear
- - - 0 - - -
flashes, as the balance calibrates a zero
setting.
---0---
The specified middle weight flashes.
Place the indicated weight in the center of the Pan.
2000
Busy
flashes. After a few seconds
Clear Pan
flashes.
Remove the mass from the Pan.
BUSY
The specified maximum weight flashes.
Place the indicated weight in the center of the Pan.
4000
Busy
flashes. After a few seconds
Clear Pan
flashes.
Remove the mass from the Pan.
BUSY
- - - 0 - - -
flashes, as the balance re-calibrates the zero
setting.
---0---
When calibration finishes,
Done
appears briefly, and the
balance returns to weighing mode.
Done
Note:
Be careful not to touch the scale or the table while calibration is in progress,
as it will cause the process to fail.
Содержание Pioneer PA64
Страница 1: ...SERVICE MANUAL Ohaus Corporation 19A Chapin Road P O Box 2033 Pine Brook NJ 07058 2033 973 377 9000 ...
Страница 2: ......
Страница 10: ...TABLE OF CONTENTS Pioneer Balances Service Manual viii Ohaus Corporation www ohaus com ...
Страница 106: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 18 Pioneer Balances Service Manual ...
Страница 115: ......