CHAPTER 3 MAINTENANCE PROCEDURES
Pioneer Balances Service Manual
3-31
Ohaus Corporation www.ohaus.com
3.3.18 Reassembling Analytical Load Cells
13. Install the Hanger: insert two screws loosely. Distance shown by arrows should be
8mm. Insert shims on each side and tighten screws. (See Figures 3-86 and 3-87.)
Figure 3-86. Screws from Bottom Mounting
Plate inserted finger-tight into the Hanger.
Figure 3-87. Insert shims between the Hanger
and the Load Cell on both sides.
Figure 3-88. Bottom Flexure Arm.
Figure 3-89. Top Flexure Arm.
14. Install the Bottom Flexure Arm,
screw-side down
. Each screw has two
washers: a silver-colored one near the Flexure, a black one near the
screw. (See Figure 3-88.)
15. Install the Top Flexure Arm,
screw-side down
. (See Figure 3-89.)
16. Install the Vertical Flexure, bottom screw last. Figure 90. Vertical Flexure.
Note the order of parts, as shown in Figure 3-91.
Figure 3-91. Vertical Flexure and screws, Lock Plate,
washers, and bushings. Note the different-sized
screws and bushings for top and bottom. Also note the
elongated hole on the bottom side of the Vertical Flexure.
Top Bushing
Top Screw
Lock Plate
Elongated hole
on bottom
Bottom Screw
Содержание Pioneer PA64