EN-60
EXPLORER BALANCES
4.9
Differential Weighing
Differential weighing stores weight values so samples can be dried or processed and the difference in weight
calculated. Up to 99 samples can be stored.
Differential weighing has two operating modes:
Auto-Process On
, where the initial and final weights are
sequentially gathered for a set of items, with no need to
return to Item Settings.
Auto-Process Off
, where initial and final weights are
gathered separately for each item, after each item is
selected in Item Settings.
Touch the
Setup
icon (wrench) to enter
Application Setup
from
the home screen.
The
DIFFERENTIAL WEIGHING
Home screen
Main Display Line
Second Display Line
Reference Fields
Application Buttons
Functions Application Icon
4.9.1
Item Settings
Press
Item Settings
to set up items in memory.
To set up the item,
Press Add New
.
New items are added with default name and weight
information (no initial or final weights).
To change the default Item Name, press
Edit Name
.
The alpha-numeric entry screen appears.