EXPLORER BALANCES
EN-29
4.2.3
Parts Counting
– Check
This feature permits establishing a set quantity of pieces as criteria for similar items that can be quickly checked
against a sample.
Press the
Setup
(wrench)
button to enter
the Counting
Setup menu.
Change
Counting
Mode to
Check
Counting
.
The
PARTS COUNTING
– Check
Home screen
Main Display Line
Second Display Line
Reference Fields
Application Buttons
Functions
The default (or last) Average Piece Weight (APW) is
displayed. Place objects on the pan to display the
Check Status
(number of pieces).
4.2.2.1 Edit Settings:
To use the Parts Counting-Check mode,
the
Over Limit
and
Under Limit
values must be set.
The
Average Piece Weight
and standard sample size can also
be set. Follow these steps:
Touch the
Edit Settings
button. The Settings screen appears.
The buttons for
APW
,
Samples
,
Over Limit
and
Under Limit
show default (or previously set) values. To change any value,
touch the button. A keyboard screen appears.
(See keyboard screens, next page.)
To set a new average piece weight, press the
Recalculate APW
button.
A new screen appears with the message:
Place 30 pieces. Press Accept
(Default is 10 pieces, but message matches
number of pieces in
Samples
button.)
After following these instructions, press
Return to Application
on the main
Edit Settings
screen.