EXPLORER BALANCES
EN-25
The Settings screen is now displayed.
Settings available:
APW, Sample Size,
Functions available:
Recalculate APW,
Return to Application
Establish an Average Piece Weight (APW)
Each time a new type of part is counted, the nominal weight of
one piece (Average Piece Weight or APW) must be established
using a small quantity of pieces. This APW is stored until
replaced by another APW.
There are two methods to establish the APW value:
1. The actual APW is known; or
2. The APW must be derived by weight. For this case the
current sample size will be used
To choose a different value, establish it beforehand, using the
Sample Size
button.
(See below.)
Set a
known
Average Piece Weight (APW)
To adjust the APW value directly, touch the
APW
button
.
A numeric input window appears.
Key in the desired APW Weight, then press
.
The display returns to the previous screen with the
new APW value highlighted.
To return to the Counting home screen,
touch
Return to Application
.
Set a new Average Piece Weight (APW)
– Derived
To establish a new APW, touch the
Recalculate APW
button.
The APW Screen is displayed.
Note:
The sample size that is displayed will be used.
To use a different sample size, change that first.
(See below.)