41
MENUS
Custom Unit or Volume Selection (Cont.)
EXPONENTS
E-3
Moves decimal point 3
places to the left.
E-2
Moves decimal point 2
places to the left.
E-1
Moves decimal point 1 place
to the left.
E0
Leaves decimal point in
normal position.
E1
Moves decimal point 1 place
to the right.
E2
Moves decimal point 2
places to the right.
E3
Moves decimal point 3
places to the right.
LSD’s
LSD .5*
Adds one decimal
place display counts by 5’s.
LSD 1
Display counts by 1’s.
LSD 2
Display counts by 2’s.
LSD 5
Display counts by 5’s.
LSD 10
Display counts by 10’s.
LSD
100 Display counts by
100’s.
* Sensitivity to vibration is in-
creased with this LSD setting.
5.
ON
TARE
. next digit flashes.
6. Repeat steps 4 and 5, and set value of all digits.
7.
to backup for errors.
8. After the last digit is entered, the display indicates the current exponent preceded by
the letter
. There are 7 exponent value which you can choose
from (see table).
9.
MODE
to change the exponent.
10.
ON
TARE
. When released, the display shows the current least significant digit. The
least significant digit is the digit in the last decimal place on the display. The selection
you make causes the balance to count by 1’s, 2’s or 5’s in this position. There are 6
1. Access the
submenu under the Setup Options
menu.
2.
ON
TARE
.
3.
ON
TARE
. The
mantissa of the current conversion is
displayed. The mantissa of the current
conversion factor is displayed. This is a
number between 0.1 and 1.999999 with
the first digit flashing. For conversion
factors outside of this range, the expo-
nent will be used to move the decimal
point.
4.
MODE
changes first digit.
LSD settings you can choose from (see
table).
11.
MODE
to change the LSD.
12.
ON
TARE
.
13.
MODE
. Den-
sity permits the selection of the density of
a liquid by measuring the volume by
weight. If the Factor is the density of a
liquid, the appropriate unit of volume can
be selected for printing.
UNITS
UNITS
UNITS
UNITS
facTOR
DeNsITy
e 0
1.000000
cuNIT
facTOR
1.000000
Menu
Menu