Safety
In order to comply with the requirements of safety standard EN 61010, the
viscometer has a separable cord and plug, without a locking device, to mate
with a socket outlet in the building. If the user substitutes another cord, the
cord should not exceed 3 meters (9'10") in length.
Any interruption of the protective earth conductor inside or outside the
instrument, or disconnection of the protective earth terminal is likely to
make the instrument dangerous under some fault conditions. Intentional
interruption is prohibited.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment might be impaired. Whenever it
is likely that protection has been impaired, the instrument should be made
inoperative, and secured against accidental operation. OFITE Technical
Service should be contacted for advice.
Any service performed by removing the viscometer cover should only be
performed by a knowledgeable operator.
There are no user-serviceable components inside the heat cup cover. The
Heat Cup should only be serviced by a factory-authorized technician.