4. Click OK to connect to the viscometer.
5. Press the Enter key twice to bring up the prompt.
6. Type M1 and press Enter.
The characters you type will not appear on the screen.
7. Type C and press Enter.
8. Ignore the instructions on the screen. Slowly turn the bob until it hits one
of the stops. Then slowly turn it in the other direction until it hits the other
stop. Then let go of the bob and let it return to its resting position.
9. Type N and press Enter. The software will show details of the calibration.
Close Hyperterminal.