Patent No. 6,766,028 - OFI Testing Equipment, Inc.
OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
43
19. Place the bearing guide back into the body and secure it in place with the
three standoffs.
20. Reattach the encoder board to the bearing guide.
21. Carefully place the spring mounting plate onto the three standoffs.
Secure it in place with the three hex screws.
Be careful not to bend the torsion spring.
22. Locate the notch on the transducer body. Turn the transducer
counterclockwise all the way against the stop. Adjust the position of the
notch on the transducer so that it rests on the far right edge of the black
mark on the encoder board.
23. Tighten the set screw on the zero calibration knob. This will lock the
spring in place.
24. Test the transducer to make sure it turns freely and returns to the
designated resting place.
25. Place the housing cover back onto the unit and secure it in place with the
five screws.
26. Adjust the P value (see page 49 for instructions). Then recalibrate the
viscometer (see page 14).