6-6. Calibration Command
Explanation of Remote Calibration command.
CALibration:VOLTage {voltage|MIN|MAX}
Voltage Calibration Command.
Please remind Calibration sequence
Please remind Calibration sequence.
"Refer 4-6 Calibration(for GPIB) using REMOTE INTERFACE"
> voltage Save MAX & MIN range calibration value.
> MIN
Panel Low range voltage calibration.
> MAX
Panel High range voltage calibration.
CAL:VOLT MIN
Operate Low Value
CAL:VOLT MIN
Operate Low Value.
CAL:VOLT voltage
Send measured Voltage value from DVM.
CAL:VOLT MAX
Operate High Value.
CAL:VOLT voltage
Sned measured Current value from DVM.
From Low to High, Power Supply saves calibration Data into
fixed memory..
CALibration:CURRent {current|MIN|MAX}
CALibration:CURRent {current|MIN|MAX}
Command related to current calibration.
Remind sequence of Calibration.
Read reference "4-6 Calibration(for GPIB) using REMOTE INTERFACE
> current Saves measured calibrated voltage from Minimum to Maximum range.
> MIN
Panel Low range current calibration.
> MAX
Panel High range current calibration
> MAX
Panel High range current calibration.
ex) Operate sequence of current calibration.
CAL:CURR MIN
Operated low value.
CAL:CURR current
Send current value to DAM.
CAL:CURR MAX
Operate High value when Low value is done.
CAL:CURR current
Send current value to DAM.
- 70 -