3-6. I/O Config & LOCAL
To configure the power supply for the RS-232 and RS485 interface, proceed as follows.
RS232C interface is standard, in case of RS485, it is optional and remove the RS232C module.
If interface is RS485 when it is shipped, address default is no. 05.
Baud rate default is 9600bps and setting of RS232C & RS485 is able to set on front panel.
▌Th RS 232 b
d
t
d
it
l
ti
t
d i
l til
▌The RS-232 baud rate and parity selections are stored in non-volatile memory,
and does not change when power has been off or after a remote interface reset.
▌If remote interface works, the lamp of RMT on front panel lits and power supply is
remotely controlled preferentially.
▌If you want to control power supply on local mode, at first finish remote nterface
and then press the "I/O & LOCAL" key. Lamp of "RMT" turn off and you can contorol.
RS232C Setting
Sequence of RS232C Setting.
▌Turn on the power supply
▌Turn on the power supply
After turned on, check the power supply displays "**OUTPUT OFF**".
▌Press "I/O" key in other to set RS232C interface.
LCD Display
▌RS232C or GPIB will be displayed following previous condition.
If "
I/O - GPIB
" displayed, press cursor key once more to "
I/O - RS232C
"
is displayed
I/O - RS232C
I/O - GPIB
is displayed.
When VFD Displays
press cursor key
and change into
If previous Display description is "
I/O - RS232C
" , perss I/O CONFIG Key.
VFD Display Description
I/O - RS232C
BAUD-RATE 4800
I/O - GPIB
▌Baud rate is divided into 4800bps, 9600bps, 19200bps, 38400bps.
It must be the fitted with PC Interface and Baud rate.
For instance, to select 19200bps, press right cursor key twice.
VFD Display Description
▌To save changes, press I/O CONFIG Key
BAUD-RATE 19200
▌To save changes, press I/O CONFIG Key.
VFD Display Description
Now, Interface would be RS232C and Baud rate is 19200bps.
What is bps? The initial of Bit per Second, data transmission unit of 1bit per one second.
CHANGE SAVED