7. Other meter functions

Automatic shut-off mode

If no key is pressed during a 20 minute period, the meter automatically shuts off to
conserve power. To restart the 20 minute count, turn the meter on again.

To disengage the automatic shut-off mode for a given power-on cycle:

17

16

7.1

1. 

Press OFF.

2. 

Press Mn/Mx and ON simultaneously
and hold until the screen displays
“nSL”, indicating that it is in the non-
sleep mode. Release the keys.

See figure

NOTE: 

To switch back to the automatic
shut-off mode, press OFF then ON.
The meter will enter automatic
shut-off mode each time you turn it
on unless you perform the above
steps to disengage this feature.

Q

T

T

4. 

Press PRG 

to save the value for 

this alarm set point. The screen 
displays “SA” then the current value 
for the set point.

See figure 

5. 

To reset another alarm set point, repeat
steps 2 through 4. Otherwise, press RST
to return to the measurement mode.

To cancel an alarm setting:

1. 

In the alarm set mode, select the alarm
set point you wish to cancel.

i)  In the measurement mode, press PRG.

ii)  Select the alarm set point you wish to

change.

a)  Press NEXT, if necessary, until the

code (Lo Temp, Hi Temp, Lo %RH,
or Hi %RH) corresponding to the set
point you wish to change is flashing.

b)  Press PRG. The screen displays the

set point you selected to change,
such as “Lo Temp”, and the current
value for this alarm set point.

2. 

Press NEXT and release. Press the RST
key and hold for two seconds.

A beep

sounds when you press the key, and two
more beeps sound after two seconds.
Release the key.

See figure

3. 

Press RST

to return to measurement

mode.

To cancel all four alarm settings at once:

1. 

Press OFF.

2. 

Press RST and ON simultaneously
and hold until the screen displays 
“OAL” and two beeps sound.

See figure

3. 

Release the keys. All alarm settings 
are now cancelled and you are in the 
measurement mode.

S

S

F

LOTEMP
HITEMP
LO%RH

HI%RH

R

R

Q

Содержание 35612-series

Страница 1: ...mer misuse Out of warranty products will be repaired on a charge basis 13 Return of items Authorization must be obtained from our Customer Service Department before returning items for any reason When...

Страница 2: ...mode 17 7 2 Min Max memory mode 18 7 3 Hold mode 19 7 4 C F mode 19 8 Advanced Operations 20 9 Troubleshooting Error Messages 21 10 Specifications 22 11 Accessories 23 12 Warranty 24 13 Return of Item...

Страница 3: ...e next digit for alarm setting Displays the next alarm setting value Selects the next digit for calibration PRG Switches from measuring program to alarm check mode Switches from alarm check to alarm s...

Страница 4: ...The probe cable has a notched 6 pin connector to attach the probe to the meter NOTE Do not substitute other probes For a replacement probe see the Accessories section page 23 NOTE Keep connector dry...

Страница 5: ...new temperature offset value Tnew by adding the results in steps a and b and subtracting the results in c Tnew To Tm Ts Tnew __________ Other Connections Three other connections are available Serial...

Страница 6: ...es not meet your accuracy requirements If the values are only slightly different and you do not require the highest possible accuracy then calibration is not necessary Two point relative humidity cali...

Страница 7: ...nt Measuring dewpoint model WD 35612 10 only I J Fast humidity calibration Follow the directions below to calibrate humidity more quickly Please note that using this calibration method may result in a...

Страница 8: ...re value in C press Mn Mx to place or remove a sign in front of the old value ii Press NEXT if necessary one or more times until the digit you need to change is flashing See figure iii Press Mn Mx unt...

Страница 9: ...s 2 through 4 Otherwise press RST to return to the measurement mode To cancel an alarm setting 1 In the alarm set mode select the alarm set point you wish to cancel i In the measurement mode press PRG...

Страница 10: ...ays SA for saved The screen then displays RH See figure 6 Press OFF When the meter is turned on again it will default to the new temperature measuring mode X X Min Max memory mode Press Mn Mx to displ...

Страница 11: ...Indicates Cause Solution Readout flashing OP Beeping Probe incorrectly Improper setup Check probe connected or faulty wire connection 0 0 Probe is Rough handling or Replace the 99 9 damaged defective...

Страница 12: ...Adapter 110 VAC WD 35615 08 AC Adapter 220 VAC 10 Specifications Mode Relative Dew Point Temperature Humidity WD 35612 10 only Range 0 to 99 9 RH 51 0 to 49 0 C 0 to 50 0 C 60 0 to 120 0 F 32 0 to 12...

Отзывы: