Hold mode

1. 

Press HOLD

while in the measurement

mode to freeze the display screen 
readings.

See figure

2. 

Press HOLD again to return to active
readings.

°C/°F mode

Toggling between °C and °F

To toggle between °C and °F: 

1. 

Press C/F.

2. 

Press again to toggle to previous mode.   

7.3

7.4

F

HOLD

%RH

W

W

Changing the default display to °C or °F

Whenever the meter is turned on, it reverts to the default
display (°C or °F). To change the default display:

1. 

Press OFF

.

2. 

Press ON and PRG simultaneously.
Hold until the screen displays
“CAL”, then release the keys.

See figure

3. 

Press RST.

The screen displays the

current default temperature unit,
“°F” or “°C”.

4. 

Press Mn/Mx

to change from the

current default temperature unit to
the desired default.

5. 

Press PRG

and hold until the screen

displays “SA”, for “saved”. The
screen then displays %RH.

See figure

6. 

Press OFF.

When the meter is turned on again it
will default to the new temperature
measuring mode.

X

X

Min/Max memory mode

Press Mn/Mx to display the minimum and maximum values of temperature and 
relative humidity that were registered during the current power-on usage period.

7.2

1. 

In the measuring mode, press
Mn/Mx. 

The display shows the 

minimum values for relative 
humidity and temperature.
“MIN”displays in the upper 
left-hand display area. If no key 
is pressed within 10 seconds, the 
display reverts to active readings.

See figure

2. 

To display the maximum values,
press Mn/Mx again

within the 

10-second period. The display shows
the maximum values for relative
humidity and temperature. “MAX”
in the upper left-hand display area.
If no key is pressed within 10 sec-
onds, the display reverts to active
readings. Or, press Mn/Mx to return
to active readings immediately.

See figure

F

MIN

%RH

U

U

F

MAX

%RH

V

V

Y

Y

Содержание 35612-series

Страница 1: ...mer misuse Out of warranty products will be repaired on a charge basis 13 Return of items Authorization must be obtained from our Customer Service Department before returning items for any reason When...

Страница 2: ...mode 17 7 2 Min Max memory mode 18 7 3 Hold mode 19 7 4 C F mode 19 8 Advanced Operations 20 9 Troubleshooting Error Messages 21 10 Specifications 22 11 Accessories 23 12 Warranty 24 13 Return of Item...

Страница 3: ...e next digit for alarm setting Displays the next alarm setting value Selects the next digit for calibration PRG Switches from measuring program to alarm check mode Switches from alarm check to alarm s...

Страница 4: ...The probe cable has a notched 6 pin connector to attach the probe to the meter NOTE Do not substitute other probes For a replacement probe see the Accessories section page 23 NOTE Keep connector dry...

Страница 5: ...new temperature offset value Tnew by adding the results in steps a and b and subtracting the results in c Tnew To Tm Ts Tnew __________ Other Connections Three other connections are available Serial...

Страница 6: ...es not meet your accuracy requirements If the values are only slightly different and you do not require the highest possible accuracy then calibration is not necessary Two point relative humidity cali...

Страница 7: ...nt Measuring dewpoint model WD 35612 10 only I J Fast humidity calibration Follow the directions below to calibrate humidity more quickly Please note that using this calibration method may result in a...

Страница 8: ...re value in C press Mn Mx to place or remove a sign in front of the old value ii Press NEXT if necessary one or more times until the digit you need to change is flashing See figure iii Press Mn Mx unt...

Страница 9: ...s 2 through 4 Otherwise press RST to return to the measurement mode To cancel an alarm setting 1 In the alarm set mode select the alarm set point you wish to cancel i In the measurement mode press PRG...

Страница 10: ...ays SA for saved The screen then displays RH See figure 6 Press OFF When the meter is turned on again it will default to the new temperature measuring mode X X Min Max memory mode Press Mn Mx to displ...

Страница 11: ...Indicates Cause Solution Readout flashing OP Beeping Probe incorrectly Improper setup Check probe connected or faulty wire connection 0 0 Probe is Rough handling or Replace the 99 9 damaged defective...

Страница 12: ...Adapter 110 VAC WD 35615 08 AC Adapter 220 VAC 10 Specifications Mode Relative Dew Point Temperature Humidity WD 35612 10 only Range 0 to 99 9 RH 51 0 to 49 0 C 0 to 50 0 C 60 0 to 120 0 F 32 0 to 12...

Отзывы: