Oakton 35612-series Скачать руководство пользователя страница 5

The optional AC adapter gives
you the flexibility of operating
from an AC power source in your
work station or laboratory. The
AC adapter is not included with
your meter; order separately on
page 23.

1.

Insert the AC jack as shown in
figure at right.

2.

Switch off the meter before
plugging the adapter into the
power source. 
This safety precaution protects
the software in your meter.

3. 

Press the ON/OFF button to
switch meter on.

4. Calibration

Temperature Calibration

NOTE:  

You don't need to calibrate the temperature unless there is a significant 
difference between the indicated meter temperature and a verified 
temperature standard.

Calibrate the temperature as follows:

1. 

Determine the amount of adjustment needed:

i)  Determine the temperature offset value currently used by 

the meter, To, by pressing OFF, then ON. The offset value 

displays for about 2 seconds just after the test screen has 

To = __________

appeared and before the current readings are shown.

ii)  Measure the temperature with this meter, Tm.

Tm = __________ 

iii)  Measure the temperature with a verified 

temperature standard, Ts.

Ts = __________

iv)  Calculate the new temperature offset value, Tnew, by 

adding the results in steps (a) and (b) and subtracting 
the results in (c):  Tnew = To + Tm – Ts.

Tnew = __________

Other Connections 

Three other connections are 
available:

•  Serial output for temperature 

and humidity

•  Analog output for humidity
•  Analog output for temperature

Make these connections according to
figure at right. See page 20 for an
explanation of how these connec-
tions may be used.

8

9

3.3

3.4

4.1

Bottom view with cover removed, 

showing output connections.

Front of meter

Serial

humidity

output

(TTL)

Analog

humidity

output

Analog

temperature

outlet

Connecting the Optional AC Adapter 

2. 

Press OFF

.

3. 

Press ON and PRG simultaneously
and hold until the screen displays
“CAL”. Release the keys. The screen dis-
plays the current temperature offset
value, To, with the first digit flashing.

See figure 

4. 

Change the temperature offset value to
the new temperature offset value, Tnew,

calculated above in step 1 (iv).

i)  Press Mn/Mx to change the flashing

digit to the first digit of Tnew.  

The number will change each time
Mn/Mx is pressed from 0 to 1, 2, 3,
–3, –2, –1, –0.

ii)  Press NEXT. The other digit flashes.

iii)  Press Mn/Mx to change flashing

digit to the second digit of Tnew.

5. 

Press PRG 

and hold until the screen

displays “SA”, for “saved”.

6. 

Press OFF.

The new temperature offset

value will now be used.

A

A

Содержание 35612-series

Страница 1: ...mer misuse Out of warranty products will be repaired on a charge basis 13 Return of items Authorization must be obtained from our Customer Service Department before returning items for any reason When...

Страница 2: ...mode 17 7 2 Min Max memory mode 18 7 3 Hold mode 19 7 4 C F mode 19 8 Advanced Operations 20 9 Troubleshooting Error Messages 21 10 Specifications 22 11 Accessories 23 12 Warranty 24 13 Return of Item...

Страница 3: ...e next digit for alarm setting Displays the next alarm setting value Selects the next digit for calibration PRG Switches from measuring program to alarm check mode Switches from alarm check to alarm s...

Страница 4: ...The probe cable has a notched 6 pin connector to attach the probe to the meter NOTE Do not substitute other probes For a replacement probe see the Accessories section page 23 NOTE Keep connector dry...

Страница 5: ...new temperature offset value Tnew by adding the results in steps a and b and subtracting the results in c Tnew To Tm Ts Tnew __________ Other Connections Three other connections are available Serial...

Страница 6: ...es not meet your accuracy requirements If the values are only slightly different and you do not require the highest possible accuracy then calibration is not necessary Two point relative humidity cali...

Страница 7: ...nt Measuring dewpoint model WD 35612 10 only I J Fast humidity calibration Follow the directions below to calibrate humidity more quickly Please note that using this calibration method may result in a...

Страница 8: ...re value in C press Mn Mx to place or remove a sign in front of the old value ii Press NEXT if necessary one or more times until the digit you need to change is flashing See figure iii Press Mn Mx unt...

Страница 9: ...s 2 through 4 Otherwise press RST to return to the measurement mode To cancel an alarm setting 1 In the alarm set mode select the alarm set point you wish to cancel i In the measurement mode press PRG...

Страница 10: ...ays SA for saved The screen then displays RH See figure 6 Press OFF When the meter is turned on again it will default to the new temperature measuring mode X X Min Max memory mode Press Mn Mx to displ...

Страница 11: ...Indicates Cause Solution Readout flashing OP Beeping Probe incorrectly Improper setup Check probe connected or faulty wire connection 0 0 Probe is Rough handling or Replace the 99 9 damaged defective...

Страница 12: ...Adapter 110 VAC WD 35615 08 AC Adapter 220 VAC 10 Specifications Mode Relative Dew Point Temperature Humidity WD 35612 10 only Range 0 to 99 9 RH 51 0 to 49 0 C 0 to 50 0 C 60 0 to 120 0 F 32 0 to 12...

Отзывы: