Oakton 35612-series Скачать руководство пользователя страница 6

i)  Gently open the cover of the 33% RH

calibration salt bottle and carefully
insert the probe into the bottle.
The probe will hit a stopper when 
it is fully inserted.

ii)  Press RST twice. The meter will beep

each time the key is pressed. Do not
press any other keys! 
The screen dis-
plays a percent humidity and “MIN”.
Low humidity calibration begins
automatically and takes 20 to 45 
minutes to complete.

See figure

iii)  When low point calibration is com-

plete, the meter will beep once and 
display “SA”, “MIN”, and “MAX”.

See figure 

5.  

Calibrate the high humidity point:

i)  Gently open the cover of the 

75% RH calibration salt bottle and
carefully insert the probe into the
bottle. 

The probe will hit a stopper

when it is fully inserted.

ii)  Press NEXT.  The screen displays 

a percent humidity and “MAX”.
High humidity calibration begins
automatically and takes 20 to 45 
minutes to complete.

See figure

iii)  When high point calibration is 

complete, the meter beeps once 
and displays “SA” briefly. Meter
automatically returns to measure-
ment mode.

Relative Humidity Calibration

10

11

MIN

%RH

4.2

B

C

MIN

MAX

D

Before calibration

Check the relative humidity calibration of your meter every 30 days or more
often if you frequently measure extreme humidity conditions. To do this, place
the probe in each calibration salt bottle and check the measured value against
the reference value. 

• If the measured value does not equal the reference value, recalibration 

may be necessary (if the difference between the values does not meet your
accuracy requirements). 

• If the values are only slightly different and you do not require the highest

possible accuracy, then calibration is not necessary.

Two-point relative humidity calibration is necessary to maintain the accuracy
of the meter throughout its full range. Relative humidity calibration takes 
20 to 45 minutes per point.

IMPORTANT: 

You will not get good results if your calibration temperature

varies more than ±0.5°F in five minutes. If this happens, the meter automatical-
ly extends the calibration time, so if the temperature is very unstable the 
calibration might never be complete. 

To prevent this, thermally isolate the calibration salt bottle and probe, or place
them in a temperature-controlled environment, such as the carrying case. Keep
away from sunlight, electronic equipment, drafts, heating ducts, and all other
causes of temperature variation.

Automatic Humidity calibration:

1. 

Place the meter, probe, and calibration
salt bottles in a place where the 
temperature will remain constant 
(< ±0.5°F) for 40 to 90 minutes. 

2. 

Press OFF.

3. 

Press ON and PRG simultaneously
and hold until the screen displays
“CAL”. 

Release the keys. The screen

displays the temperature offset value.

See figure

4. 

Calibrate the low humidity point:

CAUTION: 

Be careful when inserting

the sensor probe into the calibration salt
bottles. Rough movement may damage
the bottle.

MAX

%RH

E

E

B

D

C

Notes on automatic humidity calibration

While the meter is calibrating, do not press any key. If you do, calibration will stop and
you will have to start the process over again. 
If you need to discontinue the calibration
process, press RST twice and the meter will return to the measurement mode.

Содержание 35612-series

Страница 1: ...mer misuse Out of warranty products will be repaired on a charge basis 13 Return of items Authorization must be obtained from our Customer Service Department before returning items for any reason When...

Страница 2: ...mode 17 7 2 Min Max memory mode 18 7 3 Hold mode 19 7 4 C F mode 19 8 Advanced Operations 20 9 Troubleshooting Error Messages 21 10 Specifications 22 11 Accessories 23 12 Warranty 24 13 Return of Item...

Страница 3: ...e next digit for alarm setting Displays the next alarm setting value Selects the next digit for calibration PRG Switches from measuring program to alarm check mode Switches from alarm check to alarm s...

Страница 4: ...The probe cable has a notched 6 pin connector to attach the probe to the meter NOTE Do not substitute other probes For a replacement probe see the Accessories section page 23 NOTE Keep connector dry...

Страница 5: ...new temperature offset value Tnew by adding the results in steps a and b and subtracting the results in c Tnew To Tm Ts Tnew __________ Other Connections Three other connections are available Serial...

Страница 6: ...es not meet your accuracy requirements If the values are only slightly different and you do not require the highest possible accuracy then calibration is not necessary Two point relative humidity cali...

Страница 7: ...nt Measuring dewpoint model WD 35612 10 only I J Fast humidity calibration Follow the directions below to calibrate humidity more quickly Please note that using this calibration method may result in a...

Страница 8: ...re value in C press Mn Mx to place or remove a sign in front of the old value ii Press NEXT if necessary one or more times until the digit you need to change is flashing See figure iii Press Mn Mx unt...

Страница 9: ...s 2 through 4 Otherwise press RST to return to the measurement mode To cancel an alarm setting 1 In the alarm set mode select the alarm set point you wish to cancel i In the measurement mode press PRG...

Страница 10: ...ays SA for saved The screen then displays RH See figure 6 Press OFF When the meter is turned on again it will default to the new temperature measuring mode X X Min Max memory mode Press Mn Mx to displ...

Страница 11: ...Indicates Cause Solution Readout flashing OP Beeping Probe incorrectly Improper setup Check probe connected or faulty wire connection 0 0 Probe is Rough handling or Replace the 99 9 damaged defective...

Страница 12: ...Adapter 110 VAC WD 35615 08 AC Adapter 220 VAC 10 Specifications Mode Relative Dew Point Temperature Humidity WD 35612 10 only Range 0 to 99 9 RH 51 0 to 49 0 C 0 to 50 0 C 60 0 to 120 0 F 32 0 to 12...

Отзывы: