[F4 X] refers to dynamic testing-time.
X=0.5 / 1.0 / 2.0 / 3.0 (s). The time for scale from dynamic to stabilization depends on the dynamic
testing-time.
If you need to change the above data, press [→] to select and then press [ENTER] to confirm. If
don’t, directly press [ENTER], at this time the indicator enters F5 setup.
[F5 X] refers to digital wave-filtering strength.
X=0 / 1 / 2 / 3. The digital wave-filtering is used for eliminating the weighing change caused by
environment like quake or noise. The lower the wave-filtering strength, the quicker the indicator
response-speed, and the less stably the indicator works. On the other hand, the higher the
wave-filtering strength, the slower the indicator response-speed, and the more stably the indicator
works. The data to be selected are as follows:
0——no wave-filtering
1——low wave-filtering
2——medium wave-filtering
3——high wave-filtering
If you need to change the above data, press [→] to select and then press [ENTER] to confirm. If
don’t, directly press [ENTER], at this time the indicator enters F6 setup.
[F6 X] refers to serial output mode, there are 2 mode to be selected.
X=0 / 1.
X=0 refers to printing output (printer model: TPμP-16S).
X=1 refers to continuous output.