then press [ENTER] to confirm, at this time the indicator displays “PASS”(the indicator is reading
the range datum of the scale ). After several seconds, the indicator automatically turns to the
ambient data group, thus the range calibrating is finished.
Step 4: ambient data group
[F1 X] refers to power-on auto-zero range.
“X=0” refers to no auto-zero.
“X=10” allow to auto-zero,and the zero range is ±10% Max. weighing.
“X=20” allow to auto-zero,and the zero range is ±20% Max. weighing.
If you need to change the above data, press [→] to select and then press [ENTER] to confirm. If
don’t, directly press [ENTER], at this time the indicator enters F2 setup.
[F2 X] refers to auto zero-tracking range
X=0.5 / 1.0 / 2.0/3.0 (d). The auto zero-tracking function can auto compensate the zero-point
change caused by the factor that a little material to be weighed drop down the platform when
weighing or due to the temperature change.
If you need to change the above data, press [→] to select and then press [ENTER] to confirm. If
don’t, directly press [ENTER], at this time the indicator enters F3 setup.
[F3 X] refers to dynamic testing-range.
X=0.5 / 1.0 / 2.0 / 3.0 (d). The non-dynamic condition when to be up for the scale depends on the
dynamic testing-range. Never process the zeroing or deducting-tare operation while the scale is
under dynamic.
If you need to change the above data, press [→] to select and then press [ENTER] to confirm. If
don’t, directly press [ENTER], at this time the indicator enters F4 setup.