15
600 ㎏
0.02kg; 0.05 ㎏; 0.1 ㎏; 0.2 ㎏;
300 ㎏
1000 ㎏
0.1 ㎏; 0.2 ㎏; 0.5 ㎏; 1 ㎏
500 ㎏
2000 ㎏
0.1kg; 0.2 ㎏; 0.5 ㎏; 1 ㎏;
1000 ㎏
3000 ㎏
0.1kg; 0.2 ㎏; 0.5 ㎏; 1 ㎏;
1000 ㎏
5000 ㎏
0.2kg 0.5 ㎏; 1 ㎏; 2 ㎏;
2000 ㎏
Step 2: zero-load adjusting
Remove the goods on the scale platform after the indicator display “E SCL”, await 10 seconds
and then press [ENTER] to confirm. At this time the indicator automatically reads the zero datum
of the scale and enters range calibrating.
Step 3: range calibrating
If you select “CAL 1”, the indicator enters the following “a”. The indicator enters “b” when you
select “CAL 2”
a.
Fixed point calibration
Await several seconds, the indicator displays “Ad LD”. Load the weights on the scale platform
after the indicator display “Ad LD” for 2 seconds. Await 10 seconds (make sure that the
indicator is fully stable), then press [ENTER] to confirm, at this time the indicator displays
“PASS”(the indicator is reading the range datum of the scale ). After several seconds, the
indicator automatically turns to the ambient data group, thus the range calibrating is finished.
b. Arbitrary point calibration
Await several seconds, the indicator displays “Ad LD” and displays “‘0’00000” (the decimal
point place have automatically inputted according to the range ) after 2 seconds. At this time,
load the weights (the weights: 5%Wmax<Wadd<Wmax ) on the scale platform and input the
weights value for the indicator. Await 10 seconds (make sure that the indicator is fully stable),