Area
Visual
indication
Alarm type
Alarm
description
Solution
SpO
2
SpO
2
SpO
2
higher than
defined
Verify patient
condition
SpO
2
SpO
2
1-3 % below
defined
Verify patient
condition
SpO
2
SpO
2
more than
3% below defined
Verify patient
condition
SpO
2
SpO
2
lower than
80%
Verify patient
condition
Pulse
Pulse
Pulse higher than
defined
Verify patient
condition
Pulse
Pulse higher than
200 BPM
Pulse
Pulse lower than
40 BPM
Signal
No signal
No signal from
oximeter for less
than 300 secs
(adjustable)
Readjust the
sensor on the
patient’s f
inger
No signal
No signal from
oximeter for more
than 300 secs
(adjustable)
No signal
Oximeter power
failure
Restart device,
if alarm persist
O
2
matic PRO
100 needs
service
Not
connected
Not connected to
oximeter
Connect the
oximeter to
O
2
matic PRO
100. If the
problem
persists, send
dvice for
service
Содержание OMC PC-100
Страница 1: ...2022 6 09 ...
Страница 14: ...D 8 4 1 20 O2matic User Manual IFU 14 Figure 5 O2matic PRO 100 Admin menu Setting up a new profile ...
Страница 16: ...D 8 4 1 20 O2matic User Manual IFU 16 ...
Страница 18: ...D 8 4 1 20 O2matic User Manual IFU 18 ...
Страница 24: ...D 8 4 1 20 O2matic User Manual IFU 24 ...
Страница 26: ...D 8 4 1 20 O2matic User Manual IFU 26 ...
Страница 27: ...D 8 4 1 20 O2matic User Manual IFU 27 ...
Страница 28: ...D 8 4 1 20 O2matic User Manual IFU 28 ...
Страница 29: ...D 8 4 1 20 O2matic User Manual IFU 29 ...
Страница 36: ...D 8 4 1 20 O2matic User Manual IFU 36 ...
Страница 41: ...D 8 4 1 20 O2matic User Manual IFU 41 Figure 17 SpO2 outside safe range prompt ...
Страница 74: ...D 8 4 1 20 O2matic User Manual IFU 74 Notes ...
Страница 75: ...D 8 4 1 20 O2matic User Manual IFU 75 Notes ...
Страница 76: ...D 8 4 1 20 O2matic User Manual IFU 76 Notes ...