![O2matic OMC PC-100 Скачать руководство пользователя страница 23](http://html.mh-extra.com/html/o2matic/omc-pc-100/omc-pc-100_user-manual_719910023.webp)
D 8.4.1.20 O
2
matic User Manual - IFU
23
3
Starting Treatment
Turn on The O
2
matic PRO 100 by pressing
(A). After turning
on the device the Home Screen will appear.
The O
2
matic PRO 100 is designed to be as simple as possible and for
maximum safety, the device always
starts up in the “O2matic mode”
(see section 4.1). Thus, the device is already in treatment mode and
an alarm will start immediately if there is no signal from a pulse
oximeter sensor. To mute the alarm, press
(G)
, and the alarm
will be muted for 2 minutes. The button is lid to indicate that it is
muted. Entering any menu will also mute the alarms.
Upon startup the device resumes in the latest selected profile. If the
device was switched off while in manual mode, it will regardless, for
safety reasons, start up in O2matic mode.
Содержание OMC PC-100
Страница 1: ...2022 6 09 ...
Страница 14: ...D 8 4 1 20 O2matic User Manual IFU 14 Figure 5 O2matic PRO 100 Admin menu Setting up a new profile ...
Страница 16: ...D 8 4 1 20 O2matic User Manual IFU 16 ...
Страница 18: ...D 8 4 1 20 O2matic User Manual IFU 18 ...
Страница 24: ...D 8 4 1 20 O2matic User Manual IFU 24 ...
Страница 26: ...D 8 4 1 20 O2matic User Manual IFU 26 ...
Страница 27: ...D 8 4 1 20 O2matic User Manual IFU 27 ...
Страница 28: ...D 8 4 1 20 O2matic User Manual IFU 28 ...
Страница 29: ...D 8 4 1 20 O2matic User Manual IFU 29 ...
Страница 36: ...D 8 4 1 20 O2matic User Manual IFU 36 ...
Страница 41: ...D 8 4 1 20 O2matic User Manual IFU 41 Figure 17 SpO2 outside safe range prompt ...
Страница 74: ...D 8 4 1 20 O2matic User Manual IFU 74 Notes ...
Страница 75: ...D 8 4 1 20 O2matic User Manual IFU 75 Notes ...
Страница 76: ...D 8 4 1 20 O2matic User Manual IFU 76 Notes ...