D 8.4.1.20 O
2
matic User Manual - IFU
5
electromagnetic immunity of this equipment and result in improper
operation.
g)
Portable RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer
than 30 cm (12 inches) to any part of the O
2
matic PRO device,
including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
h)
O
2
matic PRO 100 shall never be used as a substitute for personal
and qualified observations by medical and nursing staff, as O
2
matic
PRO 100 only monitors oxygenation of the blood and pulse rate, and
other parameters and patient condition can deteriorate if personal
observation is inadequate.
For additional information refer to IEC 60601-1
1.4
Cautions
a.
Do not place O
2
matic PRO 100 in a corner, as it can be difficult to
connect and disconnect accessories.
b.
The accuracy of the SpO
2
measurement may be affected if the total
sensor length (including extension cables) is greater than 3 meters.
c.
Pulse oximetry sensors may have reading troubles when used on
patients with cold extremities due to reduced blood circulation. For
more information please contact your local distributor.
d.
Follow local or national recycling instructions regarding disposal of
device and accessories.
e.
In compliance with the European Directive on Waste Electrical and
Electronic Equipment (WEEE) 2012/19/EU, do not dispose of this
product as unsorted municipal waste. This device contains WEEE
materials. Contact your distributor regarding take-back or recycling of
the O
2
matic PRO 100.
Caution
Indicates that you must be careful when executing
these instructions. Not complying with these caution
directives can cause minor injuries or equipment
damage.
Содержание OMC PC-100
Страница 1: ...2022 6 09 ...
Страница 14: ...D 8 4 1 20 O2matic User Manual IFU 14 Figure 5 O2matic PRO 100 Admin menu Setting up a new profile ...
Страница 16: ...D 8 4 1 20 O2matic User Manual IFU 16 ...
Страница 18: ...D 8 4 1 20 O2matic User Manual IFU 18 ...
Страница 24: ...D 8 4 1 20 O2matic User Manual IFU 24 ...
Страница 26: ...D 8 4 1 20 O2matic User Manual IFU 26 ...
Страница 27: ...D 8 4 1 20 O2matic User Manual IFU 27 ...
Страница 28: ...D 8 4 1 20 O2matic User Manual IFU 28 ...
Страница 29: ...D 8 4 1 20 O2matic User Manual IFU 29 ...
Страница 36: ...D 8 4 1 20 O2matic User Manual IFU 36 ...
Страница 41: ...D 8 4 1 20 O2matic User Manual IFU 41 Figure 17 SpO2 outside safe range prompt ...
Страница 74: ...D 8 4 1 20 O2matic User Manual IFU 74 Notes ...
Страница 75: ...D 8 4 1 20 O2matic User Manual IFU 75 Notes ...
Страница 76: ...D 8 4 1 20 O2matic User Manual IFU 76 Notes ...