matic User Manual - IFU
58
5.3
Possible alarm events
*Battery capacity warnings are outside general alarm indication terminology as they
are indicated with red symbols but without audio.
The device alarms with a grey background in the table are related to
the most critical internal components. A failure in one of these
components triggers the same visual
“Check status”
indication
(except sensor error). Further identification can be made in the
Device Status screen in the admin menu.
5.4
Alarm priorities
All pulse oximeter sensor related alarms share the same zone in the
following prioritized showing order:
Devic
e
Sensor
error
Oximeter sensor
Device needs
service
Check
status
Acoustic alarm
Restart device,
if alarm persist
Device needs
service
Check
status
Back up circuit
Check
status
Memory full
Do full export
of patient
data, if alarm
persist reset
device to
factory
default, if
alarm persist
device needs
service
Check
status
Valve
Restart device,
if alarm persist
device needs
service
Check
status
Flowmeter
Содержание OMC PC-100
Страница 1: ...2022 6 09 ...
Страница 14: ...D 8 4 1 20 O2matic User Manual IFU 14 Figure 5 O2matic PRO 100 Admin menu Setting up a new profile ...
Страница 16: ...D 8 4 1 20 O2matic User Manual IFU 16 ...
Страница 18: ...D 8 4 1 20 O2matic User Manual IFU 18 ...
Страница 24: ...D 8 4 1 20 O2matic User Manual IFU 24 ...
Страница 26: ...D 8 4 1 20 O2matic User Manual IFU 26 ...
Страница 27: ...D 8 4 1 20 O2matic User Manual IFU 27 ...
Страница 28: ...D 8 4 1 20 O2matic User Manual IFU 28 ...
Страница 29: ...D 8 4 1 20 O2matic User Manual IFU 29 ...
Страница 36: ...D 8 4 1 20 O2matic User Manual IFU 36 ...
Страница 41: ...D 8 4 1 20 O2matic User Manual IFU 41 Figure 17 SpO2 outside safe range prompt ...
Страница 74: ...D 8 4 1 20 O2matic User Manual IFU 74 Notes ...
Страница 75: ...D 8 4 1 20 O2matic User Manual IFU 75 Notes ...
Страница 76: ...D 8 4 1 20 O2matic User Manual IFU 76 Notes ...