Figure 23 - Test in progress
When the test countdown is finished, the third and final phase,
which is the recovery phase, will commence. During this phase, the
patient should be made ready to resume normal treatment with the
previous patient profile and the following will happen:
•
The device beeps twice
•
Test log finishes recording
•
The TIME LEFT countdown is replaced with RECOVERY
When the RECOVERY countdown concludes, the device returns to
the previous active profile.
Caution
Always stay with the patient during recovery to
verify patient status upon resuming normal
Содержание OMC PC-100
Страница 1: ...2022 6 09 ...
Страница 14: ...D 8 4 1 20 O2matic User Manual IFU 14 Figure 5 O2matic PRO 100 Admin menu Setting up a new profile ...
Страница 16: ...D 8 4 1 20 O2matic User Manual IFU 16 ...
Страница 18: ...D 8 4 1 20 O2matic User Manual IFU 18 ...
Страница 24: ...D 8 4 1 20 O2matic User Manual IFU 24 ...
Страница 26: ...D 8 4 1 20 O2matic User Manual IFU 26 ...
Страница 27: ...D 8 4 1 20 O2matic User Manual IFU 27 ...
Страница 28: ...D 8 4 1 20 O2matic User Manual IFU 28 ...
Страница 29: ...D 8 4 1 20 O2matic User Manual IFU 29 ...
Страница 36: ...D 8 4 1 20 O2matic User Manual IFU 36 ...
Страница 41: ...D 8 4 1 20 O2matic User Manual IFU 41 Figure 17 SpO2 outside safe range prompt ...
Страница 74: ...D 8 4 1 20 O2matic User Manual IFU 74 Notes ...
Страница 75: ...D 8 4 1 20 O2matic User Manual IFU 75 Notes ...
Страница 76: ...D 8 4 1 20 O2matic User Manual IFU 76 Notes ...