30
Verify load cell response
Used for checking the calibration and linearity of the load cell. See section
Maintenance & Service
for more information.
Calibrate load cell response
Used when calibrating the load cell in the AUTOCLEAVER. See section
Maintenance & Service
for more information.
Utility menu
Cleaver setup
Opens a separate window. The changes made here define the mode of operation and behavior of the cleaver. See separate description
of this window.
Download PGM parameters from cleaver
Downloads a setup file from the current connected AUTOCLEAVER. This file could be requested by service technicians when contacting
a service center for help.
Upload
PGM parameters…
Used for restoring the default setting of a
specific
AUTOCLEAVER. Note that this should only be done on request by a service technician.
Uploading an improper file to the AUTOCLEAVER will impair the device or make it stop working.
Help menu
Manual
Opens the AUTOCLEAVER manual (PDF).
License
Opens the controller software license agreement.
About
Shows information about the controller software and, if connected, the AUTOCLEAVER.
Cleaver setup
The cleaver setup defines the behaviour and mode of operation of the AUTOCLEAVER. Please note that improper settings may cause
the unit to malfunction or affects its performance. Open the cleaver setup window from the Utility menu.
Setup window
Pause before cleaving (ANG mode)
Activates the angle cleave mode of operation on the AUTOCLEAVER LDF
TM
. Should only be checked (activated) when an LDA-unit
mounted. More information about angle cleaving can be found in the AUTOCLEAVER LDA
TM
manual.
Hold fiber after cleaving (LIN mode)
Activates the fiber holder adaptor plate mode of operation on the AUTOCLEAVER LDF
TM
. Should only be checked (activated) when the
AUTOCLEAVER
TM
has a fiber holder adaptor plate mounted.
Enable piezo waste eject motion
If checked (activated), the cleaver blade moves forward after cleaving and applies a bending tension to the waste fiber part to make
sure the fiber waste part does not stick to the fixed clamp block and v-groove when ejected. This function should only be enabled
when cleaving smaller diameters of fibers (up to approximately 250µm cladding). For larger fiber, this function could create excessive
wear on the diamond blade cleaver blade. On the AUTOCLEAVER
TM
this function is enabled by factory default.
Disable abortable cleaving process
By default, this box is unchecked, making it possible for the operator to stop the cleaving sequence once it has started. If checked, the
start button will not act as a stop button when the cleave sequence has started.
Number of cleaves at each blade position
Threshold level for the number of cleaves made on each position before the AUTOCLEAVER
TM
automatically raises the blade to the
next position. Default value on the AUTOCLEAVER
TM
is 1000 cleaves.