6.3.1.  Chemical Corrosion 

This type of corrosion is caused by chemical reactions. The electrolide liquid on the surface of the 
material is the main cause of the chemical reaction.  If that electrolide liquid is allowed to stay at 
the surface, corrosion occurs. First, discoloration appears and then the metal pittens. Aluminum 
easily reacts with the ionic solutions. 
 
The other causes of corrosion are as follows, 
 

  Chemical  vapors  in  the  laboratory  environment  which  dissolve  in  the  water  on  the  rotor  (in 

refrigerated centrifuges) 

  Corrosive liquids which overflow from overfilled and unsealed tubes. (the  liquids which spread 

out during centrifugation) 

  Contaminated and non-cleaned buckets, tubes and bottles. 

 

NOTE: 

If  the  centrifuged  samples  are  corrosive,  only  rinsing  with  water  is  not  sufficient.  The 
residuals dissolve in the water and humidity on the rotor and in the buckets. 

NOTE: 

Some particles may stick to the tubes, buckets and adapters. These particles crash and 
do  harm  to  anodized  surface  during  centrifugation  and  ease  the  occurrence  of 
corrosion.  

 

6.4. 

Stress Corrosion               

This type of corrosion is caused by the force of the centrifugation of the corrosive chemical which is 
already in contact  with the alloy. As the aluminum alloy contacts with the corrosive chemical, the 
stress corrosion starts. This type of corrosion is even more dangerous than the chemical corrosion 
as the effects of this corrosion are microscopic and very difficult to observe in the course of time. 
 
The corrosive material is pushed against the aluminum a

lloy by the centrifugation “g” force during 

the  centrifugation.  This  situation  causes  the  stress  corrosion  to  occur  more  quickly  than  the 
chemical corrosion does. Microscopic cracks occur under the force of the centrifugation. 
 
Every centrifugation causes the aluminum rotor to be attacked by the chemical more and more and 
eventually  micro-cracks  decrease  the  resistance  of  the  rotor  against  the  centrifugation  force. 
Fortunately,  no  crash  occurs  just  after  the  first  micro-cracks  have  appeared  as  the  rotors  are 
manufactured according to the high safety limits. 
 
The  corrosion  of  the  small  amount  of  corrosive  materials  does  not  result  in  severe  cracks  but 
weakens the mechanical resistance of the rotor in the course of time. 
 
The rotor’s places, buckets, buckets’ edges and the base of the rotor should be checked regularly. 
If needed, they must not be used until a specialist has checked them thoroughly.                 

6.5. 

CLEANING 

  Disconnect the centrifuge before cleaning. 

  There is no need of daily cleaning unless a tube breakage occurs or any liquid spills. 

 

6.6.  ELECTRICITY 

Centrifuged at high voltages are present behind the panels. These panels are electrically 
disconnecting Do not open the centrifuge 

Содержание NF 1200

Страница 1: ...0 N VE SANAY MALZEMELER MALAT VE T CARET A NF 1200 1200 R MULTI PURPOSE BENCH TOP CENTRIFUGES USER S MANUAL Z14 K25 249 Rev No 07 Rev Date 01 2016...

Страница 2: ...lity in the case where the goods are not used in accordance with their proper intent 4 The warranty may not be claimed for damages occurred during the shipment for damages resulting from improper hand...

Страница 3: ...ove the warning labels do not operate damaged instrument do not operate the instrument with a damaged cable do not move the instrument during operation In case of a problem contact your N ve agent for...

Страница 4: ...CTING THE MEMORIZED PROGRAM 13 4 5 SAFETY INTERLOCK SYSTEM 14 4 6 IMBALANCE DETECTION SYSTEM 14 4 7 MANUAL LID OPENING 14 SECTION 5 15 OPERATING PRINCIPLES 15 5 1 PREPARATION OF THE ROTOR TO RUN 15 5...

Страница 5: ...h a cooling system which provides to operate between 9 C and 40 C It ensures safe processes by means of the locking system which does not allow the centrifuge to operate if the lid is open and which d...

Страница 6: ...ge 01 99 min Hold Position Timer Step 1 min Acceleration Rate 0 Slowest 9 Fastest Braking Rate 0 Slowest 9 Fastest Power supply 230 V 50 Hz Temp range 9 C 40 C Refrigerant Liquid R134a Temp Accuracy 1...

Страница 7: ...daptors 1x50 ml G 51 019 Set of 6 adaptors 1x50 ml cocinal G 51 020 Set of 6 adaptors 1x30 38 ml G 51 021 Set of 6 adaptors 1x15 ml cocinal B 50 022 RS 100 Angle rotor 10x10 ml 12 000 rpm 14 167xg G 5...

Страница 8: ...mm Max Speed rpm Max RCF xg RA280 Swing out Rotor 4x280ml 164 4 100 3 082 Set of 4 buckets 380 ml nominal 76 Set of 4 inserts 1x280ml 65 Set of 4 inserts 1x250ml flat 62 Set of 4 inserts 1x200ml flat...

Страница 9: ...ailure The working bench should be durable to the device weight and vibration isolated Ensure that the rotor is placed correctly prior to usage According to the standard IEC 61010 2 020 anyone and any...

Страница 10: ...d motor shaft IF MENTIONED WARNINGS ARE NOT CONSIDERED N VE WILL NOT BE RESPONSIBLE FROM THEIR RESULTS SECTION 3 SYMBOLS Symbol in the operating instructions Attention general hazard area This symbol...

Страница 11: ...y of the run time from 1 min to 99 min and hold position H 03 Acceleration Brake Temperature Display Display of the acceleration and brake rates from 0 to 10 for NF 1200 R it also displays the bowl te...

Страница 12: ...Keys They increase or decrease values during programming stage and to select the program number during stand by position By pushing value increase key it is possible to see the set values during centr...

Страница 13: ...on the time display 2 flashes Enter the run time by pushing the value increase and decrease keys 16 H means hold position Push the function key 16 and see that the acceleration led turns on and the v...

Страница 14: ...tion led turns on The set speed or RCF value is displayed during the set time duration RPM or RCF screen will appear in every 3 seconds For the operations whose speeds are higher than 2 000 rpm the ro...

Страница 15: ...electronically when an unacceptable imbalance occurs In this case the brake is applied immediately The Err 1 imbalance error is displayed and the imbalance indicator turns on The lid can not be opene...

Страница 16: ...es should be at the same height to balance the load Dynamic Balance This balance states that diametrically oppositely replaced center of the gravity of tubes are symmetrical with respect to spinning a...

Страница 17: ...t the rotor fits on the drive shaft See Figure 4 Screw the shaft nut with socket wrench to the clockwise direction Make sure that shaft nut is screwed tightly but do not expose over force to the socke...

Страница 18: ...quently with the oil provided with the centrifuge Please remove the light oil from the pins and put a small amount of fresh oil every time you grease This will ensure free swinging of the buckets Most...

Страница 19: ...e chemical the stress corrosion starts This type of corrosion is even more dangerous than the chemical corrosion as the effects of this corrosion are microscopic and very difficult to observe in the c...

Страница 20: ...o the supply The electricity installation is not defective Power is supplied In case of below written failures related error codes are shown Err 2 Motor overheat Please switch off the centrifuge for a...

Страница 21: ...20 SECTION 9 ELECTRICAL CIRCUIT DIAGRAMS 9 1 ELECTRICAL CIRCUIT DIAGRAM OF NF 1200...

Страница 22: ...21 9 2 ELECTRICAL CIRCUIT DIAGRAM OF NF 1200R...

Страница 23: ...22 SECTION 10 WARNING LABEL GROUNDED SOCKET NF 800R FUSES 2x10A NF 800 FUSES 2x4A...

Отзывы: