8
ANGLE ROTOR ACCESSORIES
Rotor
Description
Capacity
Max.Tube
Dia(mm)
Max.Radius
(mm)
Max.Speed
(rpm)
Max
RCFxg
RS600
Angle Rotor
6x100ml
38
99
9,000
8,965
Set of 6 adaptors
1x50ml
30
Set of 6 adaptors 1x50 ml conical
30.5
Set of 6 adaptors
1x30/38ml
26
Set of 6 adaptors 1x15 ml conical
17
RS100
Angle Rotor
10x10ml
16
88
12,000
14,167
Set of 10
adaptors
1x6ml
13
Set of 10
adaptors
1x1.5/2ml
11
RS24
Angle Rotor
24x1.5/2ml
11
84
14,000
18,405
Adaptor
1x500/800µl
8
Adaptor
1x200 µl PCR
6.5
Adaptor
1x250/400/700
µl
6
RS30
Angle Rotor
30x2ml
11
99
13.500
20.170
Adaptor
1x500/800µl
8
Adaptor
1x200 µl PCR
6.5
Adaptor
1x250/400/700
µl
6
2.5 PRECAUTIONS AND USAGE LIMITATIONS
Do not use the device for any purpose other than the usage purpose.
Prior to first use, the user’s manual should be read and the device is only to be used by
authorized and trained personnel. Only authorized technical personnel handle the product in
case of any failure.
The working bench should be durable to the device weight and vibration isolated.
Ensure that the rotor is placed correctly prior to usage.
According to the standard IEC 61010-2-020, anyone and any hazardous materials should not
be in the 300 mm safety zone while centrifuge is running.
Do not move the device while it is running.
Do not open the lid while rotor is spinning.
Apply the manual lid opening procedure in the case of power cut or in the case of any error.
Use only the the spare parts, rotors and accessories which are supplied by NUVE.
Load the rotor according to the explanations in the user’s manual.
Start the device after ensuring the rotor is loaded correctly.
Do not use the centrifuge in areas which are in explosive danger.
Do not centrifuge the explosive, flammable, radioactive, corrosive materials and the materials
which may react with each other.
The centrifuge and the rotor are not microbiologically leak-proof. Use tubes with leak-proof
covers, if hazardous, toxic and pathogenic microorganisms are centrifuged.
Do not use corrosive materials which may be harmful for the device integrity, rotor and
accessories.