background image

13 

 

4.4.  RUNNING THE PROGRAM 

  Switch on the centrifuge, see that the on/off switch led turns on. 

  See that the control panel activates. 

  Load your samples to the rotor and close the lid. 

 

You can run the centrifuge in two different ways, 
 

  You can make a new program (see section 5.3) 

  You can select a program from the memory 

 
PRE-COOLING 
 

While  operating  the  NF  1200  R  centrifuge  at  sub-ambient  temperatures,  in  order  to  get  the 
maximum performance, without loading the rotor run the centrifuge at 1.500 rpm, at a temperature 
which  is  3°  C  lower  than  the  required  temperature  and  at  hold  position  until  you  see  that  the 
temperature inside the bowl drops to the required temperature. Then, stop the centrifuge, load your 
samples and start centrifugation. 

4.4.1. 

SELECTING THE MEMORIZED PROGRAM

 

Select  the  requested  program  number  at  stand-by  position  by  pushing  value  increase  and 
decrease keys. All values related to the selected program will be shown on the relevant displays. 
Check that all values are correct. 
 

  Load the rotor with samples by paying attention to the dynamic and static balances. 

  Close the lid, see that the lid open warning led (11) turns off. 

  Push START key (6) and see that the start led (10) turns on. 

  The rotor starts accelerating according to the set value and the acceleration led turns on. The 

set  speed  or  RCF  value  is  displayed  during  the  set  time  duration.    (RPM  or  RCF  screen  will 
appear in every 3 seconds). For the operations whose speeds are higher than 2.000 rpm, the 
rotor accelerates according to the set  acceleration value up to 2.000 rpm and then continues 
accelerating  with  the  highest  acceleration  value,  9  to  reach  the  set  speed.  As  the  speed 
reaches the se

t value, “-“ sign appears on the acc/brk/temp display (3). (The NF 800 R shows 

the bowl temperature on the acc/brk/temp display until the centrifugation ends) 

  The elapsed time is counted down as the centrifugation starts. 

 

When  the  time  display  shows  “00”,  the  program  ends,  the  breaking  led  turns  on  and the  set 
break rate is shown at acc/brk/temp display (3). The rotor brakes with the highest break rate, 9 
until it slows down to 2.000 rpm and continues braking according to the set braking value. 

  The elapsed time until the rotor stops is counted and is shown at the time display. This is the 

time that passes until the rotor stops. 

 

When  the  speed  display  shows  ”0000”,  the  start  led  (10)  turns  off,  the  “end”  expression 
appears and the lid led (9) turns on. The user is warned by an intermittently sounding alarm. 

  Push the LID button (8) to open the lid. 

 
You may leave the centrifuge at stand-by position. 

NOTE 1 : 

You may display the RPM or RCF value on the speed display during the run by 
pushing the function button. 

 

NOTE  2 :

 

You may see all set values during the run by pushing the value increase key.

 

Содержание NF 1200

Страница 1: ...0 N VE SANAY MALZEMELER MALAT VE T CARET A NF 1200 1200 R MULTI PURPOSE BENCH TOP CENTRIFUGES USER S MANUAL Z14 K25 249 Rev No 07 Rev Date 01 2016...

Страница 2: ...lity in the case where the goods are not used in accordance with their proper intent 4 The warranty may not be claimed for damages occurred during the shipment for damages resulting from improper hand...

Страница 3: ...ove the warning labels do not operate damaged instrument do not operate the instrument with a damaged cable do not move the instrument during operation In case of a problem contact your N ve agent for...

Страница 4: ...CTING THE MEMORIZED PROGRAM 13 4 5 SAFETY INTERLOCK SYSTEM 14 4 6 IMBALANCE DETECTION SYSTEM 14 4 7 MANUAL LID OPENING 14 SECTION 5 15 OPERATING PRINCIPLES 15 5 1 PREPARATION OF THE ROTOR TO RUN 15 5...

Страница 5: ...h a cooling system which provides to operate between 9 C and 40 C It ensures safe processes by means of the locking system which does not allow the centrifuge to operate if the lid is open and which d...

Страница 6: ...ge 01 99 min Hold Position Timer Step 1 min Acceleration Rate 0 Slowest 9 Fastest Braking Rate 0 Slowest 9 Fastest Power supply 230 V 50 Hz Temp range 9 C 40 C Refrigerant Liquid R134a Temp Accuracy 1...

Страница 7: ...daptors 1x50 ml G 51 019 Set of 6 adaptors 1x50 ml cocinal G 51 020 Set of 6 adaptors 1x30 38 ml G 51 021 Set of 6 adaptors 1x15 ml cocinal B 50 022 RS 100 Angle rotor 10x10 ml 12 000 rpm 14 167xg G 5...

Страница 8: ...mm Max Speed rpm Max RCF xg RA280 Swing out Rotor 4x280ml 164 4 100 3 082 Set of 4 buckets 380 ml nominal 76 Set of 4 inserts 1x280ml 65 Set of 4 inserts 1x250ml flat 62 Set of 4 inserts 1x200ml flat...

Страница 9: ...ailure The working bench should be durable to the device weight and vibration isolated Ensure that the rotor is placed correctly prior to usage According to the standard IEC 61010 2 020 anyone and any...

Страница 10: ...d motor shaft IF MENTIONED WARNINGS ARE NOT CONSIDERED N VE WILL NOT BE RESPONSIBLE FROM THEIR RESULTS SECTION 3 SYMBOLS Symbol in the operating instructions Attention general hazard area This symbol...

Страница 11: ...y of the run time from 1 min to 99 min and hold position H 03 Acceleration Brake Temperature Display Display of the acceleration and brake rates from 0 to 10 for NF 1200 R it also displays the bowl te...

Страница 12: ...Keys They increase or decrease values during programming stage and to select the program number during stand by position By pushing value increase key it is possible to see the set values during centr...

Страница 13: ...on the time display 2 flashes Enter the run time by pushing the value increase and decrease keys 16 H means hold position Push the function key 16 and see that the acceleration led turns on and the v...

Страница 14: ...tion led turns on The set speed or RCF value is displayed during the set time duration RPM or RCF screen will appear in every 3 seconds For the operations whose speeds are higher than 2 000 rpm the ro...

Страница 15: ...electronically when an unacceptable imbalance occurs In this case the brake is applied immediately The Err 1 imbalance error is displayed and the imbalance indicator turns on The lid can not be opene...

Страница 16: ...es should be at the same height to balance the load Dynamic Balance This balance states that diametrically oppositely replaced center of the gravity of tubes are symmetrical with respect to spinning a...

Страница 17: ...t the rotor fits on the drive shaft See Figure 4 Screw the shaft nut with socket wrench to the clockwise direction Make sure that shaft nut is screwed tightly but do not expose over force to the socke...

Страница 18: ...quently with the oil provided with the centrifuge Please remove the light oil from the pins and put a small amount of fresh oil every time you grease This will ensure free swinging of the buckets Most...

Страница 19: ...e chemical the stress corrosion starts This type of corrosion is even more dangerous than the chemical corrosion as the effects of this corrosion are microscopic and very difficult to observe in the c...

Страница 20: ...o the supply The electricity installation is not defective Power is supplied In case of below written failures related error codes are shown Err 2 Motor overheat Please switch off the centrifuge for a...

Страница 21: ...20 SECTION 9 ELECTRICAL CIRCUIT DIAGRAMS 9 1 ELECTRICAL CIRCUIT DIAGRAM OF NF 1200...

Страница 22: ...21 9 2 ELECTRICAL CIRCUIT DIAGRAM OF NF 1200R...

Страница 23: ...22 SECTION 10 WARNING LABEL GROUNDED SOCKET NF 800R FUSES 2x10A NF 800 FUSES 2x4A...

Отзывы: