background image

 

SECTION 1 

INTRODUCTION 

1.1  USE AND FUNCTION 

 
The  NF  1200  /  NF  1200  R  centrifuges  are  designed  to  separate  liquids  of  different  densities  by 
applying centrifugal force.  Different types of rotors,  which are swing-out and angle  rotors, can be 
fitted to the centrifuges. 
 
The microprocessor  control system  of the centrifuges enables the user to observe and to control 
the  speed  (RPM  or  RCF),  the  acceleration  and  braking  rates,  run  time,  program  number  and 
temperature of the bowl for NF 1200 R. The microprocessor control system also displays the RPM 
or RCF value during the run. 
 
The  ventilation  system  of  the    NF  1200  limits  temperature    in  the  bowl  between  the  ambiant 
tempe 3 °C and ambiant tempe 12 °C depending on the rotor type. The NF 1200 R 
is equipped with a cooling system which provides to operate between -

9 °C and + 40 ° C. 

 
It ensures safe processes by means of the locking system which does not allow the centrifuge to 
operate if the lid is open and which does not allow the rotor to spin if the lid is not closed. 
 
The lid locking system which does not allow the centrifuge to operate if the lid is open, which also 
does not allow the lid to be opened while the rotor is spinning and the possibility to open the lid by 
pushing  only  one  key  when  the  program  ends,  provide  safe  and  easy  working  conditions.  There 
are audible and visible alarms to inform the operator when the lid is open, when the program ends 
and when any error conditions occur. In case of a power failure, the lid could be opened manually 
by using a manual lid opening tool 
 

The NF series centrifuges are manufactured according to the following standards, 
EN 61010-1, EN 61010-2-020, EN 61000-6-3, EN 50419. 
 
This device is in compliance with WEEE Regulation. 

 
 
 
 
 
 
 
 
 
 
 

Содержание NF 1200

Страница 1: ...0 N VE SANAY MALZEMELER MALAT VE T CARET A NF 1200 1200 R MULTI PURPOSE BENCH TOP CENTRIFUGES USER S MANUAL Z14 K25 249 Rev No 07 Rev Date 01 2016...

Страница 2: ...lity in the case where the goods are not used in accordance with their proper intent 4 The warranty may not be claimed for damages occurred during the shipment for damages resulting from improper hand...

Страница 3: ...ove the warning labels do not operate damaged instrument do not operate the instrument with a damaged cable do not move the instrument during operation In case of a problem contact your N ve agent for...

Страница 4: ...CTING THE MEMORIZED PROGRAM 13 4 5 SAFETY INTERLOCK SYSTEM 14 4 6 IMBALANCE DETECTION SYSTEM 14 4 7 MANUAL LID OPENING 14 SECTION 5 15 OPERATING PRINCIPLES 15 5 1 PREPARATION OF THE ROTOR TO RUN 15 5...

Страница 5: ...h a cooling system which provides to operate between 9 C and 40 C It ensures safe processes by means of the locking system which does not allow the centrifuge to operate if the lid is open and which d...

Страница 6: ...ge 01 99 min Hold Position Timer Step 1 min Acceleration Rate 0 Slowest 9 Fastest Braking Rate 0 Slowest 9 Fastest Power supply 230 V 50 Hz Temp range 9 C 40 C Refrigerant Liquid R134a Temp Accuracy 1...

Страница 7: ...daptors 1x50 ml G 51 019 Set of 6 adaptors 1x50 ml cocinal G 51 020 Set of 6 adaptors 1x30 38 ml G 51 021 Set of 6 adaptors 1x15 ml cocinal B 50 022 RS 100 Angle rotor 10x10 ml 12 000 rpm 14 167xg G 5...

Страница 8: ...mm Max Speed rpm Max RCF xg RA280 Swing out Rotor 4x280ml 164 4 100 3 082 Set of 4 buckets 380 ml nominal 76 Set of 4 inserts 1x280ml 65 Set of 4 inserts 1x250ml flat 62 Set of 4 inserts 1x200ml flat...

Страница 9: ...ailure The working bench should be durable to the device weight and vibration isolated Ensure that the rotor is placed correctly prior to usage According to the standard IEC 61010 2 020 anyone and any...

Страница 10: ...d motor shaft IF MENTIONED WARNINGS ARE NOT CONSIDERED N VE WILL NOT BE RESPONSIBLE FROM THEIR RESULTS SECTION 3 SYMBOLS Symbol in the operating instructions Attention general hazard area This symbol...

Страница 11: ...y of the run time from 1 min to 99 min and hold position H 03 Acceleration Brake Temperature Display Display of the acceleration and brake rates from 0 to 10 for NF 1200 R it also displays the bowl te...

Страница 12: ...Keys They increase or decrease values during programming stage and to select the program number during stand by position By pushing value increase key it is possible to see the set values during centr...

Страница 13: ...on the time display 2 flashes Enter the run time by pushing the value increase and decrease keys 16 H means hold position Push the function key 16 and see that the acceleration led turns on and the v...

Страница 14: ...tion led turns on The set speed or RCF value is displayed during the set time duration RPM or RCF screen will appear in every 3 seconds For the operations whose speeds are higher than 2 000 rpm the ro...

Страница 15: ...electronically when an unacceptable imbalance occurs In this case the brake is applied immediately The Err 1 imbalance error is displayed and the imbalance indicator turns on The lid can not be opene...

Страница 16: ...es should be at the same height to balance the load Dynamic Balance This balance states that diametrically oppositely replaced center of the gravity of tubes are symmetrical with respect to spinning a...

Страница 17: ...t the rotor fits on the drive shaft See Figure 4 Screw the shaft nut with socket wrench to the clockwise direction Make sure that shaft nut is screwed tightly but do not expose over force to the socke...

Страница 18: ...quently with the oil provided with the centrifuge Please remove the light oil from the pins and put a small amount of fresh oil every time you grease This will ensure free swinging of the buckets Most...

Страница 19: ...e chemical the stress corrosion starts This type of corrosion is even more dangerous than the chemical corrosion as the effects of this corrosion are microscopic and very difficult to observe in the c...

Страница 20: ...o the supply The electricity installation is not defective Power is supplied In case of below written failures related error codes are shown Err 2 Motor overheat Please switch off the centrifuge for a...

Страница 21: ...20 SECTION 9 ELECTRICAL CIRCUIT DIAGRAMS 9 1 ELECTRICAL CIRCUIT DIAGRAM OF NF 1200...

Страница 22: ...21 9 2 ELECTRICAL CIRCUIT DIAGRAM OF NF 1200R...

Страница 23: ...22 SECTION 10 WARNING LABEL GROUNDED SOCKET NF 800R FUSES 2x10A NF 800 FUSES 2x4A...

Отзывы: