11.0 Electrical/Environmental Requirements
11.1 Electrical (Supply Voltage Fluctuations Not to /- 10%)
*NU-543-300E
230 VAC,
50/60 Hz,
1 Phase,
8 Amps
*NU-543-400E
230 VAC,
50/60 Hz,
1 Phase,
10 Amps
*NU-543-500E
230 VAC,
50/60 Hz,
1 Phase,
11 Amps
*NU-543-600E
230 VAC,
50/60 Hz,
1 Phase,
11 Amps
*CE Certified
11.2 Operational Performance (for indoor use only)
Environment Temperature Range:
60
F-85
F (15
C - 30
C)
Environment Humidity:
Maximum relative humidity 80% for temperatures up to 31
C decreasing
linearly to 50% relative humidity at 40
C
Environment Altitude:
6562 Feet (2000M)
11.3 Light Exposure
Standard Fluorescent Lighting @ 150 ft. candles (1614 LUX) maximum intensity.
11.4 Installation Category: 2.0
Installation category (overvoltage category) defines the level of transient overvoltage which the instrument is designed to
withstand safely. It depends on the nature of the electricity supply and its overvoltage protection means. For example,
in CAT II, which is the category used for instruments in installations supplied from a supply comparable to public mains
such as hospital and research laboratories and most industrial laboratories, the expected transient overvoltage is 2500 V
for a 230 V supply and 1500 V for a 120 V supply.
11.5 Pollution Degree: 2.0
Pollution degree describes the amount of conductive pollution present in the operating environment. Pollution degree 2
assumes that normally only non-conductive pollution such as dust occurs with the exception of occasional conductivity
caused by condensation.
11.6 Chemical Exposure
Chemical exposure should be limited to antibacterial materials used for cleaning and disinfecting. USE OF CHLORINATED
OR HALOGEN MATERIALS IN THE CABINET MAY DAMAGE STAINLESS STEEL. Chamber decontamination can be
accomplished by paraformaldehyde, vapor phased Hydrogen Peroxide or Ethylene Oxide without degradation of cabinet
materials.
11.7 EMC Performance (classified for light industrial)
Emissions:
EN61326
Immunity:
EN61326
Class A equipment is intended for use in an industrial environment.
In the documentation for the user, a statement shall be included drawing
attention to the fact that there may be potential difficulties in ensuring
electromagnetic compatibility in other environments, due to conducted
as well as radiated disturbances.
WARNING
!
Содержание LabGard NU-543-300E
Страница 7: ...OM0284 Page 7 of 72 Rev 2 July 2021...
Страница 9: ...OM0284 Page 9 of 72 Rev 2 July 2021...
Страница 10: ...OM0284 Page 10 of 72 Rev 2 July 2021...
Страница 11: ...OM0284 Page 11 of 72 Rev 2 July 2021...
Страница 12: ...OM0284 Page 12 of 72 Rev 2 July 2021...
Страница 16: ...OM0284 Page 16 of 72 Rev 2 July 2021...
Страница 21: ...OM0284 Page 21 of 72 Rev 2 July 2021 Figure 1...
Страница 22: ...OM0284 Page 22 of 72 Rev 2 July 2021...
Страница 35: ...OM0284 Page 35 of 72 Rev 2 July 2021...
Страница 67: ...OM0284 Page 67 of 72 Rev 2 July 2021...
Страница 68: ...OM0284 Page 68 of 72 Rev 2 July 2021...
Страница 69: ...OM0284 Page 69 of 72 Rev 2 July 2021...
Страница 70: ...OM0284 Page 70 of 72 Rev 2 July 2021...
Страница 71: ...OM0284 Page 71 of 72 Rev 2 July 2021...
Страница 72: ...OM0284 Page 72 of 72 Rev 2 July 2021...